ABL800 analyzer, model numbers 393-800 and 393-801.
Recall Z-0682-2019, part of FDA recall event 81547. See Radiometer America's company page.
Radiometer America Inc started this Class II device recall on Jul 3, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 30, 2023.
- Classification: Class II
- Started: Jul 3, 2018
- Status: Terminated
The analyzer software may cause a mis-match of patient demographics and test results when a certain sequence of events occur.
- Product
- ABL800 analyzer, model numbers 393-800 and 393-801.
- Quantity
- 2,131 analyzers
- Lots and codes
- Model 393-800 ABL800 Series w/o Creatinin (All serial numbers) for the following configurations: ABL800, ABL805, ABL810, ABL815, ABL820, ABL825, ABL830, and ABL835. Model 393-801 ABL800 Series with Creatinin (All serial numbers) for the following configurations: ABL817, ABL827, and ABL837.
- Distribution
- Distribution was nationwide. There was also military/government distribution. Foreign distribution was made to Canada.
- Recalling firm
- Radiometer America Inc, Brea, CA, United States
- How the firm notified
- Letter
- FDA classified
- Dec 22, 2018
Newest first, one product per recall event; search them all
- ABL90 FLEX Analyzer REF 393090 UDI:05700693930909Z-2320-2019 · Class II · Jul 10, 2019 · Terminated
- Product: ABL800 FLEX with Crea. The ABL800 with Crea is available in the following configurations: ABL817, ABL827 and…Z-0801-2019 · Class II · Jul 2, 2018 · Terminated
- ABL90 FLEX analyzers with software versions 3.3MR1 or below. The ABL90 FLEX Analyzer is a portable, automated Analyzer…Z-0010-2019 · Class II · May 25, 2018 · Terminated
- ABL800 analyzer with FLEXQ module. Device intended for in vitro testing of samples of whole blood for the parameters…Z-0231-2018 · Class II · Oct 24, 2017 · Terminated
- AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality…Z-3237-2017 · Class II · Sep 7, 2017 · Terminated
- The ABL800 Model#:AllZ-1463-2017 · Class II · Jan 26, 2017 · Terminated
- AQURE System Software Version 2.2.0 Model #: 933-599 UDI: (01)05700699335999(10)2.2.0 Catalog number: 933-599 …Z-0899-2017 · Class II · Nov 23, 2016 · Terminated
- AQURE System; Model Number: 933-599. The AQURE System manages blood gas and immunoassay analyzers.Z-0748-2016 · Class II · Dec 29, 2015 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports