Arca
Skip to content
Tools
Tools

PANTERA Proximal Humerus Fracture Plate System

Recall Z-0599-2020, part of FDA recall event 83963. See Toby Orthopaedics's company page.

Toby Orthopaedics, Inc. started this Class II device recall on Jun 5, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Sep 16, 2020.

  • Classification: Class II
  • Started: Jun 5, 2013
  • Status: Terminated

Why It Was Recalled

Class II

Potential suture clip breaking during a humerus fixation surgery.

Record

Product
PANTERA Proximal Humerus Fracture Plate System
Quantity
1700 devices
Lots and codes
1. All implanted plates in the US from 11/26/10 to 06/05/2013 with catalog numbers TO-PHP-L73, TO-PHP-R73, TO-PHP-L83, TO-PHP-R83, TO-PHP-L120, TO-PHP-R120, TO-PHP-L160, TO-PHP-R160, TO-PHP-L180, TO-PHP-R180, TO-PHP-L220, TO-PHP-R120, TO-PHP-LS, TO-PHP-RS, TO-PHP-LE, TO-PHP-RE; 2. All implanted plates in Spain and Italy from 11/26/10 to 03/26/2014, with catalog numbers TO-PHP-L73, TO-PHP-R73, TO-PHP-L83, TO-PHP-R83, TO-PHP-L120, TO-PHP-R120, TO-PHP-L160, TO-PHP-R160, TO-PHP-L180, TO-PHP-R180, TO-PHP-L220, TO-PHP-R120, TO-PHP-LS, TO-PHP-RS, TO-PHP-LE, TO-PHP-RE; 3. All implanted plates inโ€ฆ
Distribution
Arkansas, Florida, Puerto Rico, Spain, Panama, Australia, UK
Recalling firm
Toby Orthopaedics, Inc., West Miami, FL, United States
How the firm notified
E-Mail
FDA classified
Dec 4, 2019

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports