Ziehm Solo FD. Interventional fluoroscopic x-ray system
Recall Z-0574-2025, part of FDA recall event 95709. See Orthoscan's company page.
Orthoscan, Inc. started this Class II device recall on Aug 8, 2024 (fda mandated); FDA lists it as ongoing.
- Classification: Class II
- Started: Aug 8, 2024
- Status: Ongoing
- Products in this event: 7
Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
- Product
- Ziehm Solo FD. Interventional fluoroscopic x-ray system
- Quantity
- 25
- Lots and codes
- UDI: EZIEZIEHMSOLOFD1 Serial Numbers: 54301, 54772, 54773, 54774, 54792, 54793, 54794, 54795, 54796, 54797, 54798, 54851, 54852, 54855, 55020, 55021, 55114, 55115, 55116, 55117, 55118, 55119, 55120, 55121, 55212.
- Distribution
- U.S. (including Puerto Rico).
- Recalling firm
- Orthoscan, Inc., Scottsdale, AZ, United States
- How the firm notified
- Letter
- FDA classified
- Dec 2, 2024
- Ziehm Solo. Digital Mobile C-armZ-0573-2025 · Class II · Aug 8, 2024 · Ongoing
- Ziehm Vision RFD. Interventional fluoroscopic x-ray systemZ-0578-2025 · Class II · Aug 8, 2024 · Ongoing
- Ziehm Vision system, x-ray, tomography, computed.Z-0575-2025 · Class II · Aug 8, 2024 · Ongoing
- Ziehm Vision R. Iterventional fluoroscopic x-ray systemZ-0577-2025 · Class II · Aug 8, 2024 · Ongoing
- Ziehm Vision RFD 3D. Interventinoal fluoroscopic x-ray systemZ-0579-2025 · Class II · Aug 8, 2024 · Ongoing
- Ziehm Vision FD. Interventional fluoroscopic x-ray systemZ-0576-2025 · Class II · Aug 8, 2024 · Ongoing
- Mobile Mini C-arm system Part # 1000-0005. Intended to provide the physician with general fluoroscopic…Z-0260-2018 · Class II · Aug 31, 2017 · Terminated