Arca
Skip to content
Tools
Tools

The System is designed as a modular system with components that include an Overhead Tube…

Recall Z-0553-2024, part of FDA recall event 93479. See GE Hangwei Medical Systems's company page.

GE Hangwei Medical Systems, Co. LTD started this Class II device recall on Nov 7, 2023 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Nov 7, 2023
  • Status: Ongoing
  • Products in this event: 2

Why It Was Recalled

Class II

GE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension (OTS) assembly.

Record

Product
The System is designed as a modular system with components that include an Overhead Tube Suspension with tube/collimator, wallstand, Table, X-ray generator, and cleared wireless digital detectors. The list of detectors verified and validated for use with the Discovery XR656 HD system, including their specifications, are provided in the user documentation. The System generates diagnostic The components may be combined in different configurations to meet specific customer needs. In addition, upgrade configurations are available for predicate devices. The optional image pasting function enables the operator to stitch sequentially acquired radiographs into a single image. This 510(k) is to incorporate the VolumeRad advanced application that was currently available on the Discovery XR656 product onto the Discovery XR656 HD, as well as introduce a new Metal Artifact Reduction Algorithm, and an optional standalone console to take any HelixTM acquired images via DICOM (such as from a Discovery XR656 HD, Optima XR646 HD, or Optima XR240amx) and process the images independently of the system it was acquired on.
Quantity
68 units
Lots and codes
UDI 01001952781183561122100021DX3SS2200214HLSystem ID 205620DEF1; UDI 01001952781183561122100021DX3SS2200210HLSystem ID 602ABZSCOTTSTMPO; UDI 01001952780702651122060021DX3SS2200096HLSystem ID 408885TEMP1; UDI 01001952781183561122080021DX3SS2200158HLSystem ID 303827DEF1; UDI 01001952781183561122100021DX3SS2200199HLSystem ID 561514DEF1; UDI 01001952781183561123020021DX3SS2300053HLSystem ID 727300DEF; UDI 01001952780702651122100021DX3SS2200194HLSystem ID 229340DEF; UDI 01001952780702651122090021DX3SS2200173HLSystem ID 770219DEF1; UDI 01001952780702651122090021DX3SS2200177HLSystem ID 770219DEF2;…
Distribution
US Nationwide distribution.
Recalling firm
GE Hangwei Medical Systems, Co. LTD, Beijing, N/A, China
How the firm notified
Letter
FDA classified
Dec 14, 2023

Other Products in This Recall

Recall event 93479

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.