Arobella Medical 6mm Qurette used with the Arobella Medical AR1000 Wound Therapy System.…
Recall Z-0534-2013, part of FDA recall event 63808. See Arobella Medical's company page.
Arobella Medical, LLC started this Class II device recall on Nov 5, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 4, 2015.
- Classification: Class II
- Started: Nov 5, 2012
- Status: Terminated
This recall has been initiated because the Arobella Medical 6mm Qurette Q07 Lot Numbers 0001 and 0002 can fracture during use and may release a fragment or fragments. The 6 mm Qurette is a component of the AR1000 Quostic Wound Therapy System.
- Product
- Arobella Medical 6mm Qurette used with the Arobella Medical AR1000 Wound Therapy System. Major components of the AR1000 Wound Therapy System include an ultrasonic generator, converter hand piece, done-shaped applicator tip (qurette), and a saline supply reservoir. Product Usage: The Arobella Medical AR1000 Wound Therapy System is interned for the selective dissection and fragmentation of tissue, wound debridement ( of acute and chronic wounds, burns, and diseased or necrotic tissue), and cleansing irrigation of the wound site for the removal of debris, exudates, fragments, and other matter.
- Quantity
- 99
- Lots and codes
- 0001 and 0002
- Distribution
- Worldwide distribution - US (nationwide) and the states of AL, CA, ID, IN, IA, MI, MN, MS, NJ, NM, OH, PA. SC, TX, UT, VA, WI. including the country of SAUDI ARABIA.
- Recalling firm
- Arobella Medical, LLC, Minnetonka, MN, United States
- How the firm notified
- Letter
- FDA classified
- Dec 17, 2012
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports