Integra Disposable, Convenience Kit Single Shot Epidural Tray Reorder Number 340269 Rx…

Recall Z-0533-2014, part of FDA recall event 66885. See Integra LifeSciences Corp. d.b.a. Integra Pain Management's company page.

Integra LifeSciences Corp. d.b.a. Integra Pain Management started this Class II device recall on Oct 28, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 29, 2014.

  • Classification: Class II
  • Started: Oct 28, 2013
  • Status: Terminated

Why It Was Recalled

Class II

Integra LifeSciences Corporation is recalling certain lots of Integra Kits for Epidural Anesthesia Administration as a result of a Product Advisory Notice from BD Medical for the LOR syringes indicating the plunger of the syringe may stall or stick when travelling within the syringe.

Record

Product
Integra Disposable, Convenience Kit Single Shot Epidural Tray Reorder Number 340269 Rx Only Product Usage: The Integra kits which contain a BD LOR syringe are intended to provide licensed medical professionals with a kit containing drugs and other medical devices that are used to support the administration of general, local or spinal anesthetics. The BD LOR syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement is in the epidural space by use of the Loss of Resistance Technique as detailed in medical textbooks and medical journal articles. The LOR Syringe is not intended for injection or aspiration.
Quantity
359,250
Lots and codes
Kit Cat No: 10-2433,13-2595,13-2616,13-2628,3104195,3403061,3403150,3403316,3403579,3403727,3403776,3403813,3403989,3404131,3404269,3404293,3404417,3404468,3404578,3503559,CUS025,CUS1121-01,CUS1164-02,CUS1202-04,CUS1298-03,CUS1376,CUS1376,CUS1400-01,CUS1406-03,CUS1415-04,CUS1568-01,CUS1664-01,CUS1696,CUS1723,CUS1760,CUS1773-01,CUS1797,CUS279-02,CUS416,CUS597,CUS628-02,CUS755,PISLL7,RSM394,RSM395,SS1246,SS2036-01,
Distribution
USA Nationwide Distribution and one foreign consignee in Mursalat.
Recalling firm
Integra LifeSciences Corp. d.b.a. Integra Pain Management, Salt Lake City, UT, United States
How the firm notified
Letter
FDA classified
Dec 17, 2013