Intubation Flexible Fiberscope
Recall Z-0526-2014, part of FDA recall event 66862. See Karl Storz Endoscopy-America's company page.
Karl Storz Endoscopy America Inc started this Class II device recall on Nov 13, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 27, 2014.
- Classification: Class II
- Started: Nov 13, 2013
- Status: Terminated
- Products in this event: 2
KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.
- Product
- Intubation Flexible Fiberscope, Models: 11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1, 11302BDD1, and 11302BDD2. " Transoral intubation " Transnasal intubation " Evaluation of placement of endotracheal and double lumen endotracheal tubes " Provide visual access to the larynx and tracheobronchial tree.
- Quantity
- 1516 units
- Lots and codes
- C1140, C1140E, C1160, C1160E, QKC1750, QKC1750E, QKC1694, QKC1694E
- Distribution
- Nationwide Distribution.
- Recalling firm
- Karl Storz Endoscopy America Inc, Southbridge, MA, United States
- How the firm notified
- Letter
- FDA classified
- Dec 17, 2013
- Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the…Z-0527-2014 · Class II · Nov 13, 2013 · Terminated
- Axillent Debakey Grasping Forceps, Part No. 33821DYD In general the surgeons use 33821DYD Debakey forceps to grasp…Z-0465-2014 · Class II · Nov 6, 2013 · Terminated
- 27023WU Balloon Catheter, 3 Fr., 2/pkg, sterile, contains latex. The RUSCH ureter occlusion catheter is indicated…Z-1728-2013 · Class II · May 31, 2013 · Terminated
- KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System, Instruction Manuals: PDD Telescopes, Tricam SL II, and PDD…Z-1216-2013 · Class II · Aug 16, 2012 · Terminated
- Loop, Cutting, Angled 22FR (26055G/6). Resectoscope accessories (ie. electrode, cutting loops and cold knives).Z-1316-2013 · Class II · Jun 1, 2009 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports