Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part…
Recall Z-0356-2027, part of FDA recall event 99959. See CORIN Medical's company page.
CORIN MEDICAL, LTD. started this Class II device recall on Oct 2, 2026 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Oct 2, 2026
- Status: Ongoing
Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.
- Product
- Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.
- Quantity
- 22 units
- Lots and codes
- Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.
- Recalling firm
- CORIN MEDICAL, LTD., Cirencester, United Kingdom
- How the firm notified
- Letter
- FDA classified
- Nov 3, 2026