Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part…

Recall Z-0356-2027, part of FDA recall event 99959. See CORIN Medical's company page.

CORIN MEDICAL, LTD. started this Class II device recall on Oct 2, 2026 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Oct 2, 2026
  • Status: Ongoing

Why It Was Recalled

Class II

Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.

Record

Product
Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.
Quantity
22 units
Lots and codes
Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.
Recalling firm
CORIN MEDICAL, LTD., Cirencester, United Kingdom
How the firm notified
Letter
FDA classified
Nov 3, 2026