Grab n Go Opti series VIPR system Model # PRX-9615 - Product Usage: The Grab n Go Opti…
Recall Z-0339-2020, part of FDA recall event 83857. See Western/Scott Fetzer's company page.
Western / Scott Fetzer Company started this Class II device recall on Aug 15, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Nov 18, 2020.
- Classification: Class II
- Started: Aug 15, 2019
- Status: Terminated
- Products in this event: 5
Limited access to flow settings as a result of the control knob having been rotated beyond its functional range
- Product
- Grab n Go Opti series VIPR system Model # PRX-9615 - Product Usage: The Grab n Go Opti system is an integrated delivery system intended to provide supplemental oxygen, by Rx only, to patients. When administered by properly trained personnel for oxygen deficiency and resuscitation, the device is for emergency use only. For all other medical applications, the device is Rx only. The device is MR Conditional and suitable for use outside the 1,500 gauss (0.15 T) line during MR imaging for MRI systems up to 3.0 T. The device is intended for limited duration use, such as would be necessary during patient transports.
- Quantity
- 1,287 units
- Lots and codes
- All Lot manufactured from 08/31/14 to 08/07/2019
- Distribution
- Worldwide distribution US Nationwide distributions in the states of OH, NY, PA, TX, MA, MI, KS, IL, MD, MN, WA, TN, UT, CA, NJ, LA, CT, OR, ND, RI, WI, PR, FL, ON, NV, NB, OK, IN, AZ, MS, GA, IA and countries of Costa Rica, Mexico.
- Recalling firm
- Western / Scott Fetzer Company, Westlake, OH, United States
- How the firm notified
- Telephone
- FDA classified
- Nov 6, 2019
- Grab n Go Opti series VIPR system Model # PRX-9646 - Product Usage: The Grab n Go Opti system is an integrated…Z-0342-2020 · Class II · Aug 15, 2019 · Terminated
- Grab n Go Opti series VIPR system Model # PRX-9653 - Product Usage: The Grab n Go Opti system is an integrated…Z-0343-2020 · Class II · Aug 15, 2019 · Terminated
- Grab n Go Opti series VIPR system Model # PRX-9632 - Product Usage: The Grab n Go Opti system is an integrated…Z-0341-2020 · Class II · Aug 15, 2019 · Terminated
- Grab n Go Opti series VIPR system Model # PRX-9627 - Product Usage: The Grab n Go Opti system is an integrated…Z-0340-2020 · Class II · Aug 15, 2019 · Terminated
Newest first, one product per recall event; search them all
- Various OxyTOTE Portable Oxygen System regulators, Valve Integrated Pressure Regulator, one unit per package. …Z-2069-2015 · Class I · Jan 31, 2015 · Terminated
- Model RG3000 and RG4000 Series Valve Integrated Pressure Regulator (VIPR)Z-1849-2014 · Class II · Apr 7, 2014 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports