Critical Care Decontamination System (CCDS) Compatible N95 respirators - Product Usage:…
Recall Z-0309-2021, part of FDA recall event 86347. See Battelle Memorial Institute's company page.
Battelle Memorial Institute started this Class II device recall on Aug 13, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Feb 3, 2021.
- Classification: Class II
- Started: Aug 13, 2020
- Status: Terminated
Masks processed at one site were not maintained at levels of condensation during a portion of the decontamination cycle.
- Product
- Critical Care Decontamination System (CCDS) Compatible N95 respirators - Product Usage: Used in decontaminating compatible N95 respirators2 for multiple-user3 reuse by healthcare personnel (HCP)4 to prevent exposure to pathogenic biological airborne particulates when there are insufficient supplies of Filtering Facepiece Respirators (FFRs) resulting from the Coronavirus Disease 2019 (COVID-19) pandemic.
- Quantity
- 898
- Lots and codes
- All decontaminated compatible N95 respirators distributed from Providence, RI between 07/01/2020 and 08/12/2020.
- Distribution
- US Nationwide distribution including in the states of CT, MA, RI.
- Recalling firm
- Battelle Memorial Institute, Columbus, OH, United States
- How the firm notified
- FDA classified
- Oct 29, 2020
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports