Hitachi Scenaria Whole-body X-ray CT System

Recall Z-0271-2020, part of FDA recall event 84133. See Hitachi Medical Systems America's company page.

Hitachi Medical Systems America Inc started this Class II device recall on Oct 16, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 30, 2020.

  • Classification: Class II
  • Started: Oct 16, 2019
  • Status: Terminated

Why It Was Recalled

Class II

There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.

Record

Product
Hitachi Scenaria Whole-body X-ray CT System
Quantity
39
Lots and codes
Scenaria units S5002-S5014, S5017-S5044
Distribution
The devices were distributed to the following US states: CA, FL, GA, IA, IN, MD, MT, NC, NY, OH, OK, SD, TN, TX, WY, and PR. The products were distributed to the following foreign countries: Brazil and Mexico.
Recalling firm
Hitachi Medical Systems America Inc, Twinsburg, OH, United States
How the firm notified
Letter
FDA classified
Nov 1, 2019