Mobile Mini C-arm system Part # 1000-0004. Intended to provide the physician with general…
Recall Z-0259-2018, part of FDA recall event 78784. See Orthoscan's company page.
Orthoscan, Inc. started this Class II device recall on Aug 31, 2017 (fda mandated); FDA lists it as terminated, terminated Apr 21, 2021.
- Classification: Class II
- Started: Aug 31, 2017
- Status: Terminated
- Products in this event: 4
OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user
- Product
- Mobile Mini C-arm system Part # 1000-0004. Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.
- Quantity
- 2012 total
- Lots and codes
- Serial Number List: 5E1290 5H0178 5H0125 5E833 5E1164 5F0395 5E730 5E105 5H0126 5E921 5E742 5F0400 5E756 5E0940 5H0130 5F0126 5F0282 5F0412 5E763 5E0943 5H0151 5F0229 5H0273 5F0441 5E841 5E0961 5H0172 5F0230 5E0944 5F0449 5F0232 5E0981 5H0190 5F0387 5E0945 5F0451 5H0291 5E0982 5H0333 5F0445 5E0955 5F0456 5F0177 5E1009 5E0954 5F0453 5E1000 5F0468 5F0242 5E1042 5E1043 5F0457 5E1025 5F0471 5F0295 5E1051 5E1234 5F0465 5E1177 5F0482 5H0265 5E1063 5E1280 5F0610 5E1185 5F0495 5H0292 5E1080 5E908 5H0104 5E1225 5F0497 5E0938 5E1141 5F0158 5H0114 5E1249 5F0517 5E0990 5E1151 5F0408 5E1174 5E1265 5F0533…
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Poland, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, UAE, and United Kingdom.
- Recalling firm
- Orthoscan, Inc., Scottsdale, AZ, United States
- How the firm notified
- Letter
- FDA classified
- Jan 9, 2018
- Mobile Mini C-arm system Part # 1000-0004-FD. Intended to provide the physician with general fluoroscopic…Z-0261-2018 · Class II · Aug 31, 2017 · Terminated
- Mobile Mini C-arm system Part # 1000-0001. Intended to provide the physician with general fluoroscopic…Z-0258-2018 · Class II · Aug 31, 2017 · Terminated
- Mobile Mini C-arm system Part # 1000-0005. Intended to provide the physician with general fluoroscopic…Z-0260-2018 · Class II · Aug 31, 2017 · Terminated
Newest first, one product per recall event; search them all
- Ziehm Vision FD. Interventional fluoroscopic x-ray systemZ-0576-2025 · Class II · Aug 8, 2024 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports