3M Steri-Drape, Small Drape with Incise Film
Recall Z-0256-2023, part of FDA recall event 90989. See 3M Healthcare Business's company page.
3M Healthcare Business started this Class II device recall on Oct 6, 2022 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Oct 6, 2022
- Status: Ongoing
- Products in this event: 18
During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has observed an increase in reported adhesive related skin injuries for these affected lots.
- Product
- 3M Steri-Drape, Small Drape with Incise Film, REF 1021NS, general surgery drape
- Quantity
- 104,000 units
- Lots and codes
- UDI/DI (01)50707387497410, Lot numbers: 33HT9W, 33HWN8, 33J5T9, 33JL66, 33JLLW, 33JMKR, 33JN4N, 33K7Y7, 33K888
- Distribution
- Worldwide distribution.
- Recalling firm
- 3M Healthcare Business, Saint Paul, MN, United States
- How the firm notified
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- FDA classified
- Nov 18, 2022
- 3M Steri-Drape, Small Drape with Adhesive Aperture, REF 1020, general surgery drapeZ-0253-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Large Towel Drape, REF 1010, general surgery drapeZ-0250-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Medium Drape with Incise Film, REF 1060, general surgery drapeZ-0260-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Small Towel Drape, REF 1000NSD, general surgery drapeZ-0249-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Urological Drape, REF 1071, general surgery (urology) drapeZ-0264-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Large Towel Drape, REF 1010NSD, general surgery drapeZ-0252-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Medium Drape with Wide Adhesive Aperture, REF 1033, general surgery drapeZ-0259-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Medium Drape with Adhesive Aperture, REF 1030NS, general surgery drapeZ-0258-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Large Towel Drape, REF 1010NS, general surgery drapeZ-0251-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Medium Drape with Incise Film, REF 1060NS, general surgery drapeZ-0261-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Small Towel Drape, REF 1000NS, general surgery drapeZ-0248-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Small Drape with Incise Film, REF 1021, general surgery drapeZ-0255-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Medium Drape with Adhesive Aperture, REF 1030, general surgery drapeZ-0257-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Medium Drape with Incise Film and Pouch, REF 1061NS, general surgery drapeZ-0263-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Medium Drape with Incise Film and Pouch, REF 1061, general surgery drapeZ-0262-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Small Drape with Adhesive Aperture, REF 1020NS, general surgery drapeZ-0254-2023 · Class II · Oct 6, 2022 · Ongoing
- 3M Steri-Drape, Small Towel Drape, REF 1000, general surgery drapeZ-0247-2023 · Class II · Oct 6, 2022 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports