Arca
Skip to content
Tools
Tools

OneLIF Torque Limiting Adapter

Recall Z-0246-2023, part of FDA recall event 91016. See Novapproach Spine's company page.

Novapproach Spine LLC started this Class II device recall on Oct 12, 2022 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 1, 2026.

  • Classification: Class II
  • Started: Oct 12, 2022
  • Status: Terminated

Why It Was Recalled

Class II

The torque limiting adapter may not stay locked inside of the ratcheting handle once fully engaged.

Record

Product
OneLIF Torque Limiting Adapter, AO to 1/4 SQ, 2.25Nm Limit, Catalog #015-010-0000, a component of NovApproach Spine OneLIF Interbody System instrument kit, Catalog #OneLIF INS.
Quantity
6 devices
Lots and codes
Torque adapter - Lot #104713, UDI G07001501000000; Kit IDs - 504, 505, 506, 501, 101, and 105.
Distribution
US Nationwide distribution in the states of CA and NY.
Recalling firm
Novapproach Spine LLC, Alachua, FL, United States
How the firm notified
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
FDA classified
Nov 17, 2022

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports