AS-OBGYN Information System version 7.824.x
Recall Z-0118-2019, part of FDA recall event 81057. See AS Software's company page.
AS Software, Inc. started this Class II device recall on Sep 20, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 19, 2019.
- Classification: Class II
- Started: Sep 20, 2018
- Status: Terminated
Any change made to current ultrasound machine software or hardware upgrades, adding or modifying custom fields, or adding a new Ultrasound Machine, must be tested prior to the use in production environment to ensure that ultrasound measurements are crossing over correctly to the AS system
- Product
- AS-OBGYN Information System version 7.824.x
- Quantity
- 52 systems
- Lots and codes
- Software version 7.824.x
- Distribution
- US Nationwide.
- Recalling firm
- AS Software, Inc., Englewood Cliffs, NJ, United States
- How the firm notified
- FDA classified
- Oct 16, 2018
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports