Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a…
Recall Z-0101-2015, part of FDA recall event 69242. See Estill Medical Technologies's company page.
Estill Medical Technologies, Inc started this Class II device recall on Jul 25, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 21, 2015.
- Classification: Class II
- Started: Jul 25, 2014
- Status: Terminated
Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t
- Product
- Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV Fluid Infusion Warmer. The TA-UB1 battery was designed to provide sufficient power for the Thermal Angel to warm and deliver one liter of blood or IV fluid infusion to a trauma patient.
- Quantity
- 6,886
- Lots and codes
- 06-2011; 07-2011; 08-2011; 09-2011; 03-2012; 04-2012; 05-2012; 06-2012; 07-2012; 08-2012; 01-2013; 03-2013; 08-2013; 09-2013; 10-2013; 11-2013; 12-2013; 01-2014; 02-2014; 04-2014; 07-2012; 02-2014
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Serbia.
- Recalling firm
- Estill Medical Technologies, Inc, Dallas, TX, United States
- How the firm notified
- Letter
- FDA classified
- Oct 21, 2014