Arca
Skip to content
Tools
Tools

MED-810A Zemits NDPrime Laser products

Recall Z-0095-2024, part of FDA recall event 93194. See Advance-Esthetic's company page.

Advance-Esthetic LLC started this Class II device recall on Oct 9, 2023 (fda mandated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Oct 9, 2023
  • Status: Ongoing

Why It Was Recalled

Class II

Advance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling.

Record

Product
MED-810A Zemits NDPrime Laser products
Quantity
19
Lots and codes
Model MED-810A Zemits NDPrime Laser products
Distribution
US Nationwide Distribution
Recalling firm
Advance-Esthetic LLC, Fort Lauderdale, FL, United States
How the firm notified
Letter
FDA classified
Oct 12, 2023

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.