MED-810A Zemits NDPrime Laser products
Recall Z-0095-2024, part of FDA recall event 93194. See Advance-Esthetic's company page.
Advance-Esthetic LLC started this Class II device recall on Oct 9, 2023 (fda mandated); FDA lists it as ongoing.
- Classification: Class II
- Started: Oct 9, 2023
- Status: Ongoing
Advance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling.
- Product
- MED-810A Zemits NDPrime Laser products
- Quantity
- 19
- Lots and codes
- Model MED-810A Zemits NDPrime Laser products
- Distribution
- US Nationwide Distribution
- Recalling firm
- Advance-Esthetic LLC, Fort Lauderdale, FL, United States
- How the firm notified
- Letter
- FDA classified
- Oct 12, 2023
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports