Arca
Skip to content
Tools
Tools

ArjoHuntleigh Contoura C1000 series beds are designed to accommodate patients weighing up…

Recall Z-0019-2015, part of FDA recall event 67029. See Arjo Hospital Equipment's company page.

Arjo Hospital Equipment AB started this Class II device recall on Mar 17, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Feb 14, 2018.

  • Classification: Class II
  • Started: Mar 17, 2014
  • Status: Terminated

Why It Was Recalled

Class II

There is a low but existing entrapment risk for a patient being treated on the Contoura 1000 and Contoura 1080 bariatric beds if they have a BMI < 40.

Record

Product
ArjoHuntleigh Contoura C1000 series beds are designed to accommodate patients weighing up to 990 lbs. (450 kg). Model C1000 is the basic bed. Model C1080 includes a weighing system and a power outlet for an ArjoHuntleigh approved mattress pump. Product Usage: The Contoura series bariatric beds are electrically operating medical beds with a maximum patient weight of 450 kg. This range is intended for use in hospitals to facilitate management of obese patients. An optional integrated weighing system helps with assessment and management of the patient if required.
Quantity
1,695 beds (worldwide)
Lots and codes
1) Model Number: C1000; Serial Numbers (Foreign): 62265 , 399945 , 399946 , 399949 , 416912 , 416914 , 423215 , 475309 , 530882 , 554079 , 554082 , 566463 , 584837 , 586554 , 592203 , 592204 , 592205 , 592206 , 597027 , 597028 , 605859 , 646152 , 664632 , 680670 , 680699 , 693924 , 695671 , 699563 , 712231 , 714227 , 714228 , 714229 , 716809 , 720277 , 725622 , 731631 , 745781 , 745782 , 747354 , 747355 , 748161 , 748162 , 748635 , 761635 , 766430 , 766430 , 766431 , 766431 , 771194 , 771195 , 771197 , 771198 , 798571 , 816742 , 821563 , 821564 , 821704 , 821705 , 821706 , 821707 , 822714 ,…
Distribution
Worldwide Distribution - US: Nationwide in the states of (AZ, CA, CO, CT, GA, ID, IL, MA, MD, MN, NC, NE, OH, PA, RI, SD, TX, VA, WI); and the countries of Australia, Austria, Bahrain, Belgium, Canada, Denmark, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Lebanon, Mexico, Netherlands, New Zealand, Norway, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates
Recalling firm
Arjo Hospital Equipment AB, Eslov, Sweden
How the firm notified
Letter
FDA classified
Oct 6, 2014

Arjo Hospital Equipment AB's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.