Arca
Skip to content

OCE: Cultured Epithelial Autograft

FDA product code OCE. HDE. What the classification flags say about Part 820 under the QMSR.

As of Nov 2, 2026, FDA product code OCE (Cultured Epithelial Autograft) is HDE, reviewed through HDE (Humanitarian Device Exemption), with 0 linked 510(k), De Novo or BK numbers, 0 PMAs and 0 recall product records.

Classification

Product code
OCE
Device name
Cultured Epithelial Autograft
Device class
HDE
Regulation
FDA lists no classification regulation for this code
Review panel
General, Plastic Surgery
Premarket route
HDE (Humanitarian Device Exemption)
CGMP-exempt flag
No
Life-sustaining flag
Yes
Implant flag
Yes
Third-party review
Not a candidate
Summary malfunction reporting
Eligible

FDA definition. Interactive wound and burn dressing containing a live xenogeneic component. Indicated for use in patients who have deep dermal or full thickness burns comprising a total body surface area of greater than or equal to 30%. It may be used in conjunction with split-thickness autografts, or alone in patients for whom split-thickness autografts may not be an option due to the severity and extent of their burns.

Data from openFDA as of 2 November 2026. Read the record on openFDA · FDA product code classification database

Part 820 requirements the flags point to

Part 820 rules for OCE

CGMP exemptionClassification regulation, 21 CFR parts 862 to 892
Not flaggedFDA does not flag this product code as exempt.A manufacturer of a finished device of this type falls under part 820 (21 CFR 820.1(a)). Some classification regulations exempt only part of a generic device type, so read the regulation for your device.FDA: QMSR, CGMP exemptions · 21 CFR 820.35
Design and development, ISO 13485 clause 7.321 CFR 820.10(c)
Database cannot sayFDA lists the class as HDE.820.10(c) names class II, class III and listed class I devices. The database does not give one of those classes for an HDE code. Read the rule against the device's approved status.21 CFR 820.10
Traceability, ISO 13485 clause 7.5.9.221 CFR 820.10(d)
Flag points to the ruleFDA flags this code as life-supporting or life-sustaining.The flag uses the definition in 21 CFR 860.3, which FDA says explains the phrase "support or sustain life" in 820.10(d). The rule adds a condition the flag cannot carry: failure to perform when properly used must reasonably be expected to result in a significant injury. The manufacturer decides that for its device. The code is also flagged as an implant.21 CFR 820.10 · 21 CFR 860.3 · QMSR final rule, comment 49
Complaint and record content21 CFR 820.35
Rule names this classApplies to every finished-device manufacturer, exempt or not.820.35 adds content to complaint records, servicing records and UDI recording on top of ISO 13485 clause 4.2.5. FDA says CGMP exemption does not lift these records.21 CFR 820.35 · FDA: QMSR, CGMP exemptions

These flags describe the classification of a product code. They are not a determination for a specific device or manufacturer. How a manufacturer meets 21 CFR 820.10 depends on its own device, its intended use and the full text of the rule.

Clearances and recalls on record

510(k), De Novo and BK numbers linked
0
PMA numbers linked
0
Recall product records
0
Adverse event reports
Search MAUDE for OCE

Recall records count products, and one recall event can add many. Counts show activity in a category. They do not measure its safety.

Search all 7,098 product codes, or read the guide to the Part 820 rules behind these flags.

Check your quality documents against Part 820

Arca puts your SOPs and design files in a library and compares them against the Part 820 requirements for your device in a grid, one row per document and one column per requirement, with each cell cited to its source.

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.