GXN: Plate, Cranioplasty, Preformed, Non-Alterable
FDA product code GXN. Class II, classified under 21 CFR 882.5330. What the classification flags say about Part 820 under the QMSR.
As of Nov 2, 2026, FDA product code GXN (Plate, Cranioplasty, Preformed, Non-Alterable) is Class II under 21 CFR 882.5330, reviewed through 510(k), with 57 linked 510(k), De Novo or BK numbers, 0 PMAs and 11 recall product records.
Classification
- Product code
- GXN
- Device name
- Plate, Cranioplasty, Preformed, Non-Alterable
- Device class
- Class II
- Regulation
- 21 CFR 882.5330
- Review panel
- Neurology
- Premarket route
- 510(k)
- CGMP-exempt flag
- No
- Life-sustaining flag
- No
- Implant flag
- Yes
- Third-party review
- Not a candidate
- Summary malfunction reporting
- Eligible
Data from openFDA as of 2 November 2026. Read the record on openFDA · FDA product code classification database
Part 820 requirements the flags point to
Part 820 rules for GXN
- CGMP exemptionClassification regulation, 21 CFR parts 862 to 892
- Not flaggedFDA does not flag this product code as exempt.A manufacturer of a finished device of this type falls under part 820 (21 CFR 820.1(a)). Some classification regulations exempt only part of a generic device type, so read the regulation for your device.FDA: QMSR, CGMP exemptions · 21 CFR 820.35
- Design and development, ISO 13485 clause 7.321 CFR 820.10(c)
- Rule names this classRule names Class II.21 CFR 820.10(c) requires manufacturers of class II and class III devices to comply with clause 7.3 and its subclauses.21 CFR 820.10
- Traceability, ISO 13485 clause 7.5.9.221 CFR 820.10(d)
- Not flaggedFDA does not flag this code as life-supporting or life-sustaining.The code is flagged as an implant. 820.10(d) names devices that support or sustain life. The old rule, 820.65, also named devices intended for surgical implant. FDA said the scope of 820.10(d) is substantially similar to the old rule. The flag shows implant status. It does not decide whether 820.10(d) applies.21 CFR 820.10 · 21 CFR 860.3 · QMSR final rule, comment 49
- Complaint and record content21 CFR 820.35
- Rule names this classApplies to every finished-device manufacturer, exempt or not.820.35 adds content to complaint records, servicing records and UDI recording on top of ISO 13485 clause 4.2.5. FDA says CGMP exemption does not lift these records.21 CFR 820.35 · FDA: QMSR, CGMP exemptions
These flags describe the classification of a product code. They are not a determination for a specific device or manufacturer. How a manufacturer meets 21 CFR 820.10 depends on its own device, its intended use and the full text of the rule.
Clearances and recalls on record
- 510(k), De Novo and BK numbers linked
- 57 · Search these clearances
- PMA numbers linked
- 0
- Recall product records
- 11
- Adverse event reports
- Search MAUDE for GXN
Recall records count products, and one recall event can add many. Counts show activity in a category. They do not measure its safety.
Newest GXN clearances
- ClearFit Ti Ring 14mm (171750);ClearFit Ti Ring 14mm with Slot…: Longeviti Neuro Solutions, Jun 25, 2026
- 3D MAC Titanium Cranial Plate (TCP) System: The Surgeon General, Department of the Army (TSG-DA), Jun 18, 2026
- Easymade-Ti: CG Bio Co., Ltd., Apr 9, 2026
- METICULY Patient-specific titanium mesh implant: Meticuly Co., Ltd., Jan 14, 2026
- Synthes Patient Specific Implants: Synthes GmbH, Nov 12, 2025
- KLS Martin Cranial Implants - MR Conditional; K944565: KLS-Martin Micro…: KLS-Martin L.P., Nov 10, 2025
- CustomizedBone Service: Fin-Ceramica Faenza S.P.A., Mar 28, 2024
- Longeviti ClearFit OTS Cranial Implant: Longeviti Neuro Solutions, LLC, Nov 9, 2023
- PEEK Patient Specific Cranial/Craniofacial Implant(PSCI): Kontour(XiAn) Medical Technology Co., Ltd., Oct 28, 2022
- MedCAD AccuShape Titanium Patient-Specific Cranial Implant: Medcad, Aug 26, 2022
Search all 7,098 product codes, or read the guide to the Part 820 rules behind these flags.
Check your quality documents against Part 820