Arca
Skip to content
Tools
Tools

GE Healthcare Senographe Full Field Digital Mammography System

Full field digital, system, x-ray, mammographic. PMA P990066, from GE Healthcare (Waukesha, WI). Product code MUE, class 3.

As of Nov 20, 2026, P990066 is an FDA PMA held by GE Healthcare, approved on Jan 28, 2000 after 91 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 91 days
  • Supplements: 36
  • Recalls: 4
  • Cited as a predicate by: 6 devices

Pivotal study

The main clinical evidence FDA relied on, as its Summary of Safety and Effectiveness describes it
Design
Multicenter (four sites) reader study of diagnostic cohort plus added screening cancer cases, digital vs. screen-film mammography
Enrolled
641
Where
4 sites
Primary endpoint
Non-inferiority of digital vs. screen-film recall rate, sensitivity, and ROC area; recall rate is primary safety-related concern
Result
Met: recall rate 45% digital vs. 47% SFM (p<0.001); sensitivity 78% vs. 74% (p<0.01); ROC areas within 0.1 of SFM

Cited as a predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to P990066, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 17, 2011Z-1063-2012GE Healthcare has become aware of inaccurate measurements on GE magnified images provided by Senographe 2000D Mammography systems when reviewed on IHE compliant workstations. This issue may impact patient safety.Terminated
Sep 22, 2011Z-0114-2012, Z-0115-2012GE Healthcare has become aware of inaccurate measurement on GE magnified images provided by Senographe DS and Essential Mammography systems when reviewed on IHE compliant workstations. This issue may impact patient…Terminated
Feb 6, 2004Z-0282-04The mammography system's information for users lacked the technical specification for the minimum filtration permanently installed in the useful beam and the technical specification for the maximum line current.Terminated

Supplements

36 supplements to this PMA; the 15 newest
Nov 5, 2010S037Process Change - Manufacturer/Sterilizer/Packager/Supplier135 Review Track For 30-Day Notice
May 27, 2010S036Process Change - Manufacturer/Sterilizer/Packager/Supplier30-Day Notice
May 20, 2010S035Change Design/Components/Specifications/MaterialNormal 180 Day Track
Nov 24, 2009S034Change Design/Components/Specifications/MaterialReal-Time Process
Jul 10, 2009S033Change Design/Components/Specifications/MaterialReal-Time Process
Jul 30, 2008S032Labeling Change - Indications/instructions/shelf life/tradenameSpecial (Immediate Track)
Jul 16, 2008S031Process Change - Manufacturer/Sterilizer/Packager/Supplier30-Day Notice
May 27, 2008S030Real-Time Process
Aug 1, 2007S029Change Design/Components/Specifications/MaterialReal-Time Process
Apr 13, 2007S027Change Design/Components/Specifications/MaterialReal-Time Process
Apr 13, 2007S028Change Design/Components/Specifications/MaterialReal-Time Process
Jan 3, 2007S026Labeling Change - Indications/instructions/shelf life/tradenameSpecial (Immediate Track)
Oct 20, 2006S025Process Change - Manufacturer/Sterilizer/Packager/Supplier30-Day Notice
Sep 1, 2006S024Labeling Change - Indications/instructions/shelf life/tradenameSpecial (Immediate Track)
May 5, 2006S023Labeling Change - Indications/instructions/shelf life/tradenameSpecial (Immediate Track)

Other MUE clearances

The newest 510(k)s for the same product code; their median review was 209 days; see every MUE device

Record

Type
Original PMA
Received
Oct 29, 1999
Review panel
Radiology
Regulation
21 CFR 892.1715

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files