GE Healthcare Senographe Full Field Digital Mammography System
Full field digital, system, x-ray, mammographic. PMA P990066, from GE Healthcare (Waukesha, WI). Product code MUE, class 3.
As of Nov 20, 2026, P990066 is an FDA PMA held by GE Healthcare, approved on Jan 28, 2000 after 91 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 91 days
- Supplements: 36
- Recalls: 4
- Cited as a predicate by: 6 devices
- Design
- Multicenter (four sites) reader study of diagnostic cohort plus added screening cancer cases, digital vs. screen-film mammography
- Enrolled
- 641
- Where
- 4 sites
- Primary endpoint
- Non-inferiority of digital vs. screen-film recall rate, sensitivity, and ROC area; recall rate is primary safety-related concern
- Result
- Met: recall rate 45% digital vs. 47% SFM (p<0.001); sensitivity 78% vs. 74% (p<0.01); ROC areas within 0.1 of SFM
- Senographe PristinaK162268 · GE Healthcare · Nov 3, 2016
- MAMMOMAT FusionK151645 · Siemens Medical Solutions USA, Inc. · Sep 14, 2015
- Senographe Essential, Premium View I (Pvi)K131885 · GE Medical Systems SCS · Sep 26, 2013
- Giotto Image 3D, Giotto Image 3D-LK111434 · Giotto USA, LLC · Oct 27, 2011
- Senographe DS, Senographe EssentialK110798 · General Electric Company · Dec 22, 2011
- Senograph DS, Senograph EssenstialK103485 · General Electric Company · Oct 6, 2011
| Oct 17, 2011 | Z-1063-2012GE Healthcare has become aware of inaccurate measurements on GE magnified images provided by Senographe 2000D Mammography systems when reviewed on IHE compliant workstations. This issue may impact patient safety. | Terminated |
| Sep 22, 2011 | Z-0114-2012, Z-0115-2012GE Healthcare has become aware of inaccurate measurement on GE magnified images provided by Senographe DS and Essential Mammography systems when reviewed on IHE compliant workstations. This issue may impact patient… | Terminated |
| Feb 6, 2004 | Z-0282-04The mammography system's information for users lacked the technical specification for the minimum filtration permanently installed in the useful beam and the technical specification for the maximum line current. | Terminated |
| Nov 5, 2010 | S037Process Change - Manufacturer/Sterilizer/Packager/Supplier | 135 Review Track For 30-Day Notice |
| May 27, 2010 | S036Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| May 20, 2010 | S035Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Nov 24, 2009 | S034Change Design/Components/Specifications/Material | Real-Time Process |
| Jul 10, 2009 | S033Change Design/Components/Specifications/Material | Real-Time Process |
| Jul 30, 2008 | S032Labeling Change - Indications/instructions/shelf life/tradename | Special (Immediate Track) |
| Jul 16, 2008 | S031Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| May 27, 2008 | S030 | Real-Time Process |
| Aug 1, 2007 | S029Change Design/Components/Specifications/Material | Real-Time Process |
| Apr 13, 2007 | S027Change Design/Components/Specifications/Material | Real-Time Process |
| Apr 13, 2007 | S028Change Design/Components/Specifications/Material | Real-Time Process |
| Jan 3, 2007 | S026Labeling Change - Indications/instructions/shelf life/tradename | Special (Immediate Track) |
| Oct 20, 2006 | S025Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Sep 1, 2006 | S024Labeling Change - Indications/instructions/shelf life/tradename | Special (Immediate Track) |
| May 5, 2006 | S023Labeling Change - Indications/instructions/shelf life/tradename | Special (Immediate Track) |
- I-View®K254090 · Hologic, Inc. · Sep 2, 2026
- MAMMOMAT B.brilliant (11372950)K253678 · Siemens Medical Solutions USA, Inc. · Jun 18, 2026
- Bellalun 2D (VDMS-1000S)K254092 · Vieworks Co., Ltd. · May 28, 2026
- Senographe Pristina; SenoBright HD; Pristina Serena; Pristina Serena BrightK260087 · GE Medical Systems SCS · Mar 24, 2026
- HestiaK243420 · Genoray Co., Ltd. · Jul 17, 2025
- 2430TCA with Xmaru WK243849 · Rayence Co., Ltd. · Mar 12, 2025
- Vivix-MK241113 · Vieworks Co., Ltd. · Jan 16, 2025
- MAMMOMAT B.brilliantK233539 · Siemens Medical Solutions USA, Inc. · Mar 27, 2024
- Document
- Approval order (PDF) ↗
- Type
- Original PMA
- Received
- Oct 29, 1999
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1715
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files