Arca
Skip to content
Tools
Tools

Menicon Z Rigid Gas Permeable Contact Lens

Lens, contact (rigid gas permeable), extended wear. PMA P990018, from Menicon Co, Ltd. (Nagoya). Product code MWL, class 3.

As of Nov 21, 2026, P990018 is an FDA PMA held by Menicon Co, Ltd., approved on Jul 11, 2000 after 481 days, with no recalls tied to it in FDA's recall database.

  • Review time: 481 days
  • Supplements: 10
  • Recalls: 0
  • Cited as a predicate by: 1 device

What FDA approved

FDA's approval order statement

Approval for the menicon z(TM) (tisilfocon a) rigid gas permeable lens. Menicon z(TM) (tisilfocon a) spherical or aspheric lenses and non-prism ballast toric lenses are indicated for extended wear (from 1 to 7 days between removals for cleaning and disinfection of the lenses, as recommended by the eyecare practitioner) for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in non-aphakic persons with non-diseased eyes. The lens may be disinfected using a...

Pivotal study

The main clinical evidence FDA relied on, as its Summary of Safety and Effectiveness describes it
Design
Prospective, multicenter comparative trial vs. historical control (Menicon SF-PT) for up to 7 days extended wear
Enrolled
520
Where
8 sites
Primary endpoint
Equivalence to control lens in far visual acuity (20/40 or better) at 12 months; safety: slit lamp findings grade 3 or worse
Result
Met: 99.7% of eyes had 20/40 or better vs. 99.5% historical control; slit lamp grade 3 or worse 0.8% vs. 4.3%, equivalent

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Supplements

10 supplements to this PMA
Jun 16, 2023S010Process Change - Manufacturer/Sterilizer/Packager/Supplier30-Day Notice
May 1, 2023S009Change Design/Components/Specifications/MaterialReal-Time Process
Nov 16, 2022S007Change Design/Components/Specifications/MaterialNormal 180 Day Track
Aug 2, 2022S008Labeling Change - Indications/instructions/shelf life/tradenameNormal 180 Day Track
May 6, 2021S006Labeling Change - Indications/instructions/shelf life/tradenameNormal 180 Day Track No User Fee
Nov 6, 2019S005Change Design/Components/Specifications/MaterialNormal 180 Day Track
Feb 25, 2010S004Labeling Change - PASNormal 180 Day Track No User Fee
Nov 6, 2002S003Process Change - Manufacturer/Sterilizer/Packager/SupplierReal-Time Process
Jul 12, 2002S002Labeling Change - Indications/instructions/shelf life/tradenamePanel Track
Sep 6, 2000S001Postapproval Study ProtocolNormal 180 Day Track

Record

Type
Original PMA
Received
Mar 18, 1999
Review panel
Ophthalmic
Regulation
21 CFR 886.5916

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files