Menicon Z Rigid Gas Permeable Contact Lens
Lens, contact (rigid gas permeable), extended wear. PMA P990018, from Menicon Co, Ltd. (Nagoya). Product code MWL, class 3.
As of Nov 21, 2026, P990018 is an FDA PMA held by Menicon Co, Ltd., approved on Jul 11, 2000 after 481 days, with no recalls tied to it in FDA's recall database.
- Review time: 481 days
- Supplements: 10
- Recalls: 0
- Cited as a predicate by: 1 device
Approval for the menicon z(TM) (tisilfocon a) rigid gas permeable lens. Menicon z(TM) (tisilfocon a) spherical or aspheric lenses and non-prism ballast toric lenses are indicated for extended wear (from 1 to 7 days between removals for cleaning and disinfection of the lenses, as recommended by the eyecare practitioner) for the correction of refractive error (myopia, hyperopia, presbyopia and/or astigmatism) in non-aphakic persons with non-diseased eyes. The lens may be disinfected using a...
- Design
- Prospective, multicenter comparative trial vs. historical control (Menicon SF-PT) for up to 7 days extended wear
- Enrolled
- 520
- Where
- 8 sites
- Primary endpoint
- Equivalence to control lens in far visual acuity (20/40 or better) at 12 months; safety: slit lamp findings grade 3 or worse
- Result
- Met: 99.7% of eyes had 20/40 or better vs. 99.5% historical control; slit lamp grade 3 or worse 0.8% vs. 4.3%, equivalent
- Menicon Z Rigid Gas Permeable Contact LensK081443 · Menicon Co, Ltd. · Sep 23, 2008
| Jun 16, 2023 | S010Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| May 1, 2023 | S009Change Design/Components/Specifications/Material | Real-Time Process |
| Nov 16, 2022 | S007Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Aug 2, 2022 | S008Labeling Change - Indications/instructions/shelf life/tradename | Normal 180 Day Track |
| May 6, 2021 | S006Labeling Change - Indications/instructions/shelf life/tradename | Normal 180 Day Track No User Fee |
| Nov 6, 2019 | S005Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Feb 25, 2010 | S004Labeling Change - PAS | Normal 180 Day Track No User Fee |
| Nov 6, 2002 | S003Process Change - Manufacturer/Sterilizer/Packager/Supplier | Real-Time Process |
| Jul 12, 2002 | S002Labeling Change - Indications/instructions/shelf life/tradename | Panel Track |
| Sep 6, 2000 | S001Postapproval Study Protocol | Normal 180 Day Track |
- Document
- Approval order (PDF) ↗
- Type
- Original PMA
- Received
- Mar 18, 1999
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.5916
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files