Arca
Skip to content

Inter Fix Threaded Fusion Device

Intervertebral fusion device with bone graft, lumbar. PMA P970015, from Sofamor Danek (Memphis, TN). Product code MAX, class 3.

As of Nov 17, 2026, P970015 is an FDA PMA held by Sofamor Danek, approved on May 14, 1999 after 780 days, with 1 recall tied to it in FDA's recall database.

  • Approved: May 14, 1999
  • Review time: 780 days
  • Supplements: 22
  • Recalls: 1
  • Cited as a predicate by: 17 devices

What FDA Approved

FDA's approval order statement

Approvalf or the INTER FIX(TM) Threaded Fusion Device. This device is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level. INTER FIX(TM) implants are indicated to be used with autogenous bone graft and implanted via an open anterior approach.

Pivotal Study

The main clinical evidence FDA relied on, as its Summary of Safety and Effectiveness describes it
Design
Prospective, multicenter, randomized controlled IDE trial with nonrandomized arm vs. femoral ring allograft control
Where
US
Primary endpoint
Overall success at 24 months (fusion, 15-point Oswestry improvement, neurological maintenance) vs. control, equivalence delta 0.20; safety: adverse events
Result
Met: 24-mo overall success 59.7% (37/62) device vs. 48.6% (17/35) control, statistically no worse than control at delta 0.20

Cited as a Predicate

17 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to P970015, newest first; each opens FDA's recall record
StartedReasonStatus
May 27, 2005Z-1090-05The product is labeled as sterile, but this lot was not sterilized.Terminated

Supplements

22 supplements to this PMA; the 15 newest
Jul 27, 2005S031OtherNormal 180 Day Track No User Fee
Jul 22, 2005S029Change Design/Components/Specifications/MaterialReal-Time Process
Feb 4, 2005S030Change Design/Components/Specifications/MaterialReal-Time Process
Jul 19, 2004S028Process Change - Manufacturer/Sterilizer/Packager/Supplier30-Day Notice
Apr 16, 2004S026Labeling Change - Indications/instructions/shelf life/tradenameSpecial (Immediate Track)
Apr 16, 2004S027Labeling Change - Indications/instructions/shelf life/tradenameSpecial (Immediate Track)
Sep 17, 2003S025Postapproval Study ProtocolNormal 180 Day Track No User Fee
Sep 10, 2003S022Change Design/Components/Specifications/MaterialNormal 180 Day Track
Jun 24, 2003S021Location Change - Manufacturer/Sterilizer/Packager/SupplierNormal 180 Day Track No User Fee
Aug 12, 2002S020Labeling Change - Indications/instructions/shelf life/tradenameNormal 180 Day Track
Jul 19, 2001S019Labeling Change - Indications/instructions/shelf life/tradenameNormal 180 Day Track
Dec 13, 2000S017Labeling Change - Indications/instructions/shelf life/tradenameNormal 180 Day Track
Sep 28, 2000S010Change Design/Components/Specifications/MaterialNormal 180 Day Track
Aug 9, 2000S014Express GMP SupplementNormal 180 Day Track
Jul 11, 2000S013Postapproval Study ProtocolNormal 180 Day Track

Other MAX Clearances

The newest 510(k)s for the same product code; their median review was 142 days; see every MAX device

Record

Type
Original PMA
Received
Mar 25, 1997
Review panel
Orthopedic
Regulation
21 CFR 888.3080

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.