Microny Pacemakers
Pulse-generator, single chamber, sensor driven, implantable. PMA P970013, from Abbott Medical (Sylmar, CA). Product code LWO, class 3.
As of Nov 17, 2026, P970013 is an FDA PMA held by Abbott Medical, approved on Dec 21, 2000 after 1,448 days, with no recalls tied to it in FDA's recall database.
- Approved: Dec 21, 2000
- Review time: 1,448 days
- Supplements: 94
- Recalls: 0
- Cited as a predicate by: 1 device
What FDA Approved
FDA's approval order statementApproval for the microny(tm) sr+ model 2425t single chamber rate adaptive pulse generator, which is programmed with the aps ii programmer with modle 3204a software, the apsu programmer, and the model 3500/3510 programmer with model 3304 software. The deivce is indicated for the following: 1) accepted patient conditions warranting chronic cardiac pacing which include: sick sinus syndrome; chronic, symptomatic second- and third-degree av block; recurrent adams-stokes syndrome; and symtomatic bilateral bundle branch block when tachy-arrhythmia and other causes have been ruled out. 2) atrial pacing in patients with sinus node dysfunction and normal av and intraventricular conduction systems. 3) ventricular pacing in patients with significant bradycardia and: normal sinus rhythm with only rare episodes of av block or sinus arrest requiring short periods of pacing support, chronic atrial…
Pivotal Study
The main clinical evidence FDA relied on, as its Summary of Safety and Effectiveness describes it- Design
- Multicenter clinical trial (AutoCapture part) comparing device-stored data with surface ECG and 24-hour Holter
- Enrolled
- 137
- Where
- 22 sites, US, Canada
- Primary endpoint
- Ability of AutoCapture to correctly identify loss of capture and deliver a backup safety pulse to ensure 100% pacing
- Result
- Met: every ECG-recorded capture loss recognized and followed by a backup pulse (100%); 820 of 820 Holter capture losses also 100% followed by backup pulse
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Asfora Bullet Cage SystemK090048 · Medical Designs, LLC · Aug 14, 2009
Supplements
94 supplements to this PMA; the 15 newest| Jul 21, 2026 | S099Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Apr 24, 2026 | S100Change Design/Components/Specifications/Material | Real-Time Process |
| Jun 17, 2025 | S097Change Design/Components/Specifications/Material | Normal 180 Day Track |
| May 22, 2025 | S098Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Apr 1, 2025 | S094Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Feb 5, 2025 | S096Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Dec 18, 2024 | S095Change Design/Components/Specifications/Material | Real-Time Process |
| Dec 27, 2023 | S093Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Apr 14, 2023 | S092Change Design/Components/Specifications/Material | Real-Time Process |
| Sep 30, 2022 | S091Change Design/Components/Specifications/Material | Real-Time Process |
| May 17, 2022 | S090Change Design/Components/Specifications/Material | Real-Time Process |
| Jan 7, 2022 | S089Change Design/Components/Specifications/Material | Real-Time Process |
| Sep 15, 2021 | S085Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Jul 7, 2021 | S088Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Jun 28, 2021 | S087Change Design/Components/Specifications/Material | Real-Time Process |
Record
- Document
- Approval order (PDF) ↗
- Type
- Original PMA
- Received
- Jan 3, 1997
- Review panel
- Cardiovascular
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files