Arca
Skip to content

Bak Interbody Fusion System

Intervertebral fusion device with bone graft, lumbar. PMA P950002, from Zimmer Spine, Inc. (Minneapolis, MN). Product code MAX, class 3.

As of Nov 17, 2026, P950002 is an FDA PMA held by Zimmer Spine, Inc., approved on Sep 20, 1996 after 612 days, with no recalls tied to it in FDA's recall database.

  • Approved: Sep 20, 1996
  • Review time: 612 days
  • Supplements: 13
  • Recalls: 0
  • Cited as a predicate by: 33 devices

What FDA Approved

FDA's approval order statement

Approval for the bak interbody fusion system with instrumentation.

Pivotal Study

The main clinical evidence FDA relied on, as its Summary of Safety and Effectiveness describes it
Design
Multicenter IDE study (G900193) with literature controls (ALIF/PLIF), 24-month follow-up, no concurrent randomized control
Enrolled
947
Where
19 sites, US
Primary endpoint
Overall study success: fusion at involved level(s), pain improvement, maintained or improved function and muscle strength at 12 and 24 months; primary safety endpoint is complication rates
Result
Overall success 66% (391/589) at 12 months and 72% (184/254) at 24 months; no performance goal stated in text

Cited as a Predicate

33 later 510(k) summaries name this device as a predicate; the 12 newest

Supplements

13 supplements to this PMA
Jul 1, 2004S014Change Design/Components/Specifications/MaterialReal-Time Process
Jan 29, 2004S013Change Design/Components/Specifications/MaterialReal-Time Process
Oct 31, 2003S012Change Design/Components/Specifications/MaterialReal-Time Process
Jul 8, 2003S011Change Design/Components/Specifications/MaterialNormal 180 Day Track
May 1, 2002S007Change Design/Components/Specifications/MaterialNormal 180 Day Track
Apr 3, 2001S009Labeling Change - Indications/instructions/shelf life/tradenameSpecial (Immediate Track)
Dec 20, 1999S008Change Design/Components/Specifications/MaterialReal-Time Process
May 12, 1998S006Location Change - Manufacturer/Sterilizer/Packager/SupplierNormal 180 Day Track
Feb 6, 1998S005Process Change - Manufacturer/Sterilizer/Packager/SupplierReal-Time Process
Oct 15, 1997S003Change Design/Components/Specifications/MaterialNormal 180 Day Track
Jul 22, 1997S001Labeling Change - Indications/instructions/shelf life/tradenameReal-Time Process
Jun 30, 1997S004Change Design/Components/Specifications/MaterialReal-Time Process
May 9, 1997S002Change Design/Components/Specifications/MaterialReal-Time Process

Other MAX Clearances

The newest 510(k)s for the same product code; their median review was 142 days; see every MAX device

Record

Type
Original PMA, expedited
Received
Jan 17, 1995
Review panel
Orthopedic
Regulation
21 CFR 888.3080

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.