Selenia Dimensions Full Field Digital Mammography System

Digital breast tomosynthesis. PMA P080003, from Hologic, Inc. (Bedford, MA). Product code OTE, class 3.

As of Nov 20, 2026, P080003 is an FDA PMA held by Hologic, Inc., approved on Feb 11, 2011 after 1,116 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 1,116 days
  • Supplements: 10
  • Recalls: 4
  • Cited as a predicate by: 2 devices

What FDA approved

FDA's approval order statement

Approval for the selenia dimensions 3d system. This device is indicated to generate digital mammographic images that can be used for screening and diagnosis of breast cancer. The selenia dimensions (2d or 3d) system is intended for use in the same clinical applications as 2d mammography systems for screening mammograms. Specifically, the selenia dimensions system can be used to acquire 2d digital mammograms and 3d mammograms. The screening examination will consist of a 2d image set or a 2d and 3d image set. The selenia dimensions system may also be used for additional diagnostic workup of the breast.

Pivotal study

The main clinical evidence FDA relied on, as its Summary of Safety and Effectiveness describes it
Design
Retrospective, enriched multi-reader multi-case reader study (Reader Study 1: 12 readers, 312 image sets), 2D vs. 2D plus 3D
Enrolled
1,192
Where
5 sites, US
Primary endpoint
Improved ROC AUC for 2D plus 3D vs. 2D alone and/or reduced non-cancer recall rate;
Result
Met: average ROC AUC gain 0.071 (BIRADS) and 0.072 (probability of malignancy) vs. 2D alone; screening recall rate reduced from 51.5% (2D) to 12.9% (2D plus 3D)

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to P080003, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 8, 2026Z-2168-2026Complaints have been received of systems developing loose, missing, or broken internal bolts over time.Open
Aug 19, 2024Z-3248-2024Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may…Open
Jan 16, 2023Z-1144-2023C-Arm unexpected movement may cause blunt trauma should the tube arm impinge upon an individualOpen
Jan 21, 2020Z-1343-2020The recalling firm has received reports from the field of the C-arm on the Selenia Dimensions system seizing or dropping during field repairs and cleaning of the systems.Terminated

Supplements

10 supplements to this PMA
Nov 24, 2025S011Location Change - Manufacturer/Sterilizer/Packager/SupplierNormal 180 Day Track No User Fee
Jul 5, 2024S009Change Design/Components/Specifications/MaterialNormal 180 Day Track
May 29, 2024S010Location Change - Manufacturer/Sterilizer/Packager/SupplierNormal 180 Day Track No User Fee
Oct 24, 2019S008Change Design/Components/Specifications/MaterialNormal 180 Day Track
Jun 22, 2018S007Change Design/Components/Specifications/MaterialReal-Time Process
Mar 23, 2018S006Change Design/Components/Specifications/MaterialNormal 180 Day Track
May 23, 2017S005Labeling Change - Indications/instructions/shelf life/tradenameSpecial (Immediate Track)
Apr 20, 2017S004Labeling Change - Indications/instructions/shelf life/tradenameSpecial (Immediate Track)
Jul 7, 2016S003Change Design/Components/Specifications/MaterialNormal 180 Day Track
May 16, 2013S001Change Design/Components/Specifications/MaterialPanel Track

Record

Type
Original PMA
Received
Jan 22, 2008
Review panel
Radiology

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files