Lorad Digital Breast Imager
Full field digital, system, x-ray, mammographic. PMA P010025, from Hologic, Inc. (Danbury, CT). Product code MUE, class 3.
As of Nov 20, 2026, P010025 is an FDA PMA held by Hologic, Inc., approved on Mar 15, 2002 after 319 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 319 days
- Supplements: 17
- Recalls: 4
- Cited as a predicate by: 2 devices
Approval for the Lorad Digital Breast Imager. The device generates digital mammographic images that can be used for screening and diagnosis of breast cancer. It is intended for use in the same clinical applications as traditional screen-film mammographic systems.
- Design
- Reader study with enhanced cancer population; 12 radiologists read LDBI vs. screen-film mammograms (paired cases)
- Enrolled
- 200
- Where
- 4 sites
- Primary endpoint
- Diagnostic accuracy of LDBI vs. screen-film (sensitivity, specificity, recall, ROC Az); primary endpoint not explicitly labeled in text
- Result
- Specificity +1.5% to +3.7% (n.s.); sensitivity -2.1% to -7.6% (two significant); ROC Az difference not statistically significant
- Trident HD Specimen Radiography SystemK182727 · Hologic, Inc. · Jan 10, 2019
- RSM 1824C with RConsole1K162670 · DRTECH Corporation · Feb 23, 2017
| Apr 8, 2026 | Z-2167-2026Complaints have been received of systems developing loose, missing, or broken internal bolts over time. | Open |
| May 13, 2014 | Z-1700-2014Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury | Terminated |
| Sep 6, 2013 | Z-2301-2013Software: When the Selenia upon meeting the hard drive threshold of 4,000 images, may delete the most recently dated images | Terminated |
| Feb 28, 2008 | Z-1546-2008Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication. | Terminated |
| Nov 16, 2010 | S019Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Sep 7, 2010 | S020Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Mar 4, 2010 | S018Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Dec 23, 2009 | S017Change Design/Components/Specifications/Material | Real-Time Process |
| May 1, 2009 | S016Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Mar 16, 2009 | S015Change Design/Components/Specifications/Material | Real-Time Process |
| Dec 22, 2008 | S013Change Design/Components/Specifications/Material | Normal 180 Day Track |
| May 23, 2008 | S012Change Design/Components/Specifications/Material | Real-Time Process |
| Jun 1, 2007 | S010Change Design/Components/Specifications/Material | Real-Time Process |
| Jan 17, 2007 | S009Change Design/Components/Specifications/Material | Real-Time Process |
| Mar 29, 2006 | S008Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Oct 31, 2005 | S007Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Feb 18, 2005 | S005Change Design/Components/Specifications/Material | Normal 180 Day Track |
| Mar 2, 2004 | S004Change Design/Components/Specifications/Material | Real-Time Process |
| Oct 10, 2003 | S003Change Design/Components/Specifications/Material | Normal 180 Day Track |
- I-View®K254090 · Hologic, Inc. · Sep 2, 2026
- MAMMOMAT B.brilliant (11372950)K253678 · Siemens Medical Solutions USA, Inc. · Jun 18, 2026
- Bellalun 2D (VDMS-1000S)K254092 · Vieworks Co., Ltd. · May 28, 2026
- Senographe Pristina; SenoBright HD; Pristina Serena; Pristina Serena BrightK260087 · GE Medical Systems SCS · Mar 24, 2026
- HestiaK243420 · Genoray Co., Ltd. · Jul 17, 2025
- 2430TCA with Xmaru WK243849 · Rayence Co., Ltd. · Mar 12, 2025
- Vivix-MK241113 · Vieworks Co., Ltd. · Jan 16, 2025
- MAMMOMAT B.brilliantK233539 · Siemens Medical Solutions USA, Inc. · Mar 27, 2024
- Document
- Approval order (PDF) ↗
- Type
- Original PMA
- Received
- Apr 30, 2001
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1715
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files