Lorad Digital Breast Imager

Full field digital, system, x-ray, mammographic. PMA P010025, from Hologic, Inc. (Danbury, CT). Product code MUE, class 3.

As of Nov 20, 2026, P010025 is an FDA PMA held by Hologic, Inc., approved on Mar 15, 2002 after 319 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 319 days
  • Supplements: 17
  • Recalls: 4
  • Cited as a predicate by: 2 devices

What FDA approved

FDA's approval order statement

Approval for the Lorad Digital Breast Imager. The device generates digital mammographic images that can be used for screening and diagnosis of breast cancer. It is intended for use in the same clinical applications as traditional screen-film mammographic systems.

Pivotal study

The main clinical evidence FDA relied on, as its Summary of Safety and Effectiveness describes it
Design
Reader study with enhanced cancer population; 12 radiologists read LDBI vs. screen-film mammograms (paired cases)
Enrolled
200
Where
4 sites
Primary endpoint
Diagnostic accuracy of LDBI vs. screen-film (sensitivity, specificity, recall, ROC Az); primary endpoint not explicitly labeled in text
Result
Specificity +1.5% to +3.7% (n.s.); sensitivity -2.1% to -7.6% (two significant); ROC Az difference not statistically significant

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to P010025, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 8, 2026Z-2167-2026Complaints have been received of systems developing loose, missing, or broken internal bolts over time.Open
May 13, 2014Z-1700-2014Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injuryTerminated
Sep 6, 2013Z-2301-2013Software: When the Selenia upon meeting the hard drive threshold of 4,000 images, may delete the most recently dated imagesTerminated
Feb 28, 2008Z-1546-2008Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication.Terminated

Supplements

17 supplements to this PMA; the 15 newest
Nov 16, 2010S019Change Design/Components/Specifications/MaterialNormal 180 Day Track
Sep 7, 2010S020Change Design/Components/Specifications/MaterialNormal 180 Day Track
Mar 4, 2010S018Process Change - Manufacturer/Sterilizer/Packager/Supplier30-Day Notice
Dec 23, 2009S017Change Design/Components/Specifications/MaterialReal-Time Process
May 1, 2009S016Process Change - Manufacturer/Sterilizer/Packager/Supplier30-Day Notice
Mar 16, 2009S015Change Design/Components/Specifications/MaterialReal-Time Process
Dec 22, 2008S013Change Design/Components/Specifications/MaterialNormal 180 Day Track
May 23, 2008S012Change Design/Components/Specifications/MaterialReal-Time Process
Jun 1, 2007S010Change Design/Components/Specifications/MaterialReal-Time Process
Jan 17, 2007S009Change Design/Components/Specifications/MaterialReal-Time Process
Mar 29, 2006S008Change Design/Components/Specifications/MaterialNormal 180 Day Track
Oct 31, 2005S007Change Design/Components/Specifications/MaterialNormal 180 Day Track
Feb 18, 2005S005Change Design/Components/Specifications/MaterialNormal 180 Day Track
Mar 2, 2004S004Change Design/Components/Specifications/MaterialReal-Time Process
Oct 10, 2003S003Change Design/Components/Specifications/MaterialNormal 180 Day Track

Other MUE clearances

The newest 510(k)s for the same product code; their median review was 209 days; see every MUE device

Record

Type
Original PMA
Received
Apr 30, 2001
Review panel
Radiology
Regulation
21 CFR 892.1715

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files