Affinity (Tm) Cage System
Intervertebral fusion device with bone graft, lumbar. PMA P000028, from Medtronic Sofamor Danek USA, Inc. (Memphis, TN). Product code MAX, class 3.
As of Nov 17, 2026, P000028 is an FDA PMA held by Medtronic Sofamor Danek USA, Inc., approved on Jun 13, 2002 after 702 days, with no recalls tied to it in FDA's recall database.
- Approved: Jun 13, 2002
- Review time: 702 days
- Supplements: 7
- Recalls: 0
- Cited as a predicate by: 16 devices
What FDA Approved
FDA's approval order statementApproval for the affinity cage system. This device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the c2-c3 disc to the c7-t1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. Affinity implants are to be used with autogenous bone graft and implanted via an open, anterior approach.
Pivotal Study
The main clinical evidence FDA relied on, as its Summary of Safety and Effectiveness describes it- Design
- Prospective, multicenter, non-randomized, controlled equivalence trial (IDE G960201) vs. single-level autograft iliac crest fusion
- Enrolled
- 202
- Where
- US, UK
- Primary endpoint
- Overall success at 24 months (fusion, NDI pain/disability success, neurological success, no revision/removal/supplemental fixation); safety: adverse event rates
- Result
- Overall success 68% (95% HPD 60% to 74%) device vs. 61% (48% to 75%) control; device at least statistically equivalent to control
Cited as a Predicate
16 later 510(k) summaries name this device as a predicate; the 12 newest- AcifK173148 · Coligne AG · Dec 1, 2017
- Divergence Anterior Cervical Fusion SystemK141599 · Medtronic Sofamor Danek USA, Inc. · Jan 21, 2015
- Divergence Mini-Plate Anterior Cervical Fusion SystemK140417 · Medtronic Sofamor Danek, Inc. · Jul 9, 2014
- Perimeter C Spinal SystemK132584 · Medtronic Sofamor Danek USA, Inc. · Dec 4, 2013
- Stalif CK120819 · Centinel Spine, Inc. · Aug 31, 2012
- Synster Cervical Cage, Alif Cage, Plif Cage, Ptlif Cage, Tlif CageK111820 · Bm Korea Co., Ltd. · Mar 14, 2012
- Genesys Spine Interbody Fusion SystemK103034 · Genesys Spine · Dec 10, 2010
- Trabecular Metal Fusion Device Model 06-101-XXXXX, 06-102-XXXXX And 08-100-XXXXXK103033 · Zimmer Trabecular Metal Technology · Jan 10, 2011
- Avs Anchor-C Cervical Cage SystemK102606 · Stryker Spine · Apr 22, 2011
- Cornerstone Psr Cervical Interbody Fusion DeviceK100214 · Medtronic Sofamor Danek · Jun 25, 2010
- Tryptik Ca Anterior Intersomatic Cervical Cage, Models Mos-Ca 12 05-S, Mos-Ca 12 O6-SK091873 · Spineart · Sep 18, 2009
- Ldr Spine Cervical Interbody Fusion SystemK091088 · Ldr Spine USA · Jul 14, 2009
Supplements
7 supplements to this PMA| Jul 25, 2006 | S008Labeling Change - Indications/instructions/shelf life/tradename | Normal 180 Day Track No User Fee |
| Aug 24, 2004 | S006Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Jun 18, 2004 | S005Labeling Change - Indications/instructions/shelf life/tradename | Special (Immediate Track) |
| Nov 13, 2003 | S004Postapproval Study Protocol | Normal 180 Day Track No User Fee |
| Jul 2, 2003 | S003Express GMP Supplement | Normal 180 Day Track |
| Dec 3, 2002 | S002Postapproval Study Protocol | Normal 180 Day Track |
| Aug 13, 2002 | S001Change Design/Components/Specifications/Material | Real-Time Process |
Other MAX Clearances
The newest 510(k)s for the same product code; their median review was 142 days; see every MAX device- Symmetrix PX Spinal SystemK262572 · Medtronic Sofamor Danek USA, Inc. · Oct 29, 2026
- Calibrate MAX System; Calibrate NanoTec MAX System; Calibrate CCX Interbody System;…K262646 · Alphatec Spine, Inc. · Oct 21, 2026
- Scala® TlK261049 · Spineart SA · Aug 20, 2026
- TruLift Expandable SystemK261430 · Life Spine, Inc. · Jul 28, 2026
- Catalyft PL Expandable Interbody SystemK253011 · Medtronic Sofamor Danek USA, Inc. · Jul 17, 2026
- Symmetrix L Spinal SystemK261020 · Medtronic Sofamor Danek USA, Inc. · Jul 14, 2026
- Admiral ACP System; Explorer TO System; Laminoplasty System; Manta Ray TDF Spacer;…K253420 · SeaSpine Orthopedics Corporation · Jul 1, 2026
- Fule Interbody Fusion Cage SystemK253151 · Beijing Fule Science & Technology Development Co., Ltd. · Jun 15, 2026
Record
- Document
- Approval order (PDF) ↗
- Type
- Original PMA, expedited
- Received
- Jul 11, 2000
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3080
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files