Arca
Skip to content
Tools
Tools

Connect Data Management System

Glucose Dehydrogenase, Glucose. 510(k) K994433, from Abbott Diabetes Care, Inc. (Alameda, CA). Product code LFR, class 2.

As of Nov 21, 2026, K994433 is an FDA 510(k) held by Abbott Diabetes Care, Inc., substantially equivalent on May 5, 2000 after 127 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 127 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The TheraSense Inc. FreeStyleâ„¢ Connect Data Management System is intended for use in home and clinical settings to aid people with diabetes and health care professionals in the review, analysis, and evaluation of historical blood glucose test results to support an effective diabetes management program.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K994433, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Aug 30, 2005Z-1594-05Downloadable software obtained from the firm's website displays user data (including but not limited to insulin levels and pump statistics data) in a matter that may be misinterpreted and result in mistreatment of…Terminated

Record

Type
Traditional
Received
Dec 30, 1999
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1345

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files