Arca
Skip to content
Tools
Tools

Ortho Development Balanced Knee System

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K994370, from Ortho Development Corp. (Draper, UT). Product code JWH, class 2.

As of Nov 19, 2026, K994370 is an FDA 510(k) held by Ortho Development Corp., substantially equivalent on Mar 24, 2000 after 88 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 88 days
  • Recalls: 6
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Balanced Knee™ System is intended for single use in primary total knee replacement only. Indications include: - Loss of joint configuration and joint function; • Osteoarthritis of the knee joint; • Rheumatoid arthritis of the knee joint; • Post-traumatic arthritis of the knee joint; • Moderate valgus, varus, or flexion deformities.

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K994370, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 30, 2019Z-1904-2020, Z-1903-2020The locking mechanism of the tibial tray exhibited a manufacturing defect in which a uniform ridge was cut along the inner wall of the "T" and "Y" junctions, which may cause resistance when the Tibial Insert was being…Open
Jul 11, 2014Z-2254-2014, Z-2255-2014Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after…Terminated
Apr 1, 2014Z-1610-2014A femoral component containing pegs was found in a box for the pegless version.Terminated
May 29, 2009Z-1993-2009Erroneous lot number on 3 of 28 units of Balanced Knee System. Three units from Lot Number 0055245 were mixed in with the recalled lot (0054891).Terminated

Other JWH clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Dec 27, 1999
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files