HP Carevue 9000 Clinical Information System, Model M2331A
Display, Cathode-Ray Tube, Medical. 510(k) K992636, from Hewlett-Packard Co. (Andover, MA). Product code DXJ, class 2.
As of Nov 19, 2026, K992636 is an FDA 510(k) held by Hewlett-Packard Co., substantially equivalent on Aug 31, 1999 after 25 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 25 days
- Recalls: 8
- Cited as a predicate by: 4 devices
The HP M2331A CareVue 9000 is a clinical information system intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by health care providers whenever there is a need for generation of a patient record and computation of drug dosage.
- Hewlett-Packard Model 78730 Carevue 5000K922210 · Hewlett-Packard Co. · Apr 1, 1993
- Models M1175A and M1176A Component Monitoring SystK922058 · Hewlett-Packard Co. · Oct 6, 1992
- HP Viridia Component Monitoring System, Model Rev. KK990125 · Hewlett-Packard GmbH · Jan 29, 1999
- HP M2376A Device Link SystemK984194 · Hewlett-Packard Co. · Apr 30, 1999
- Signal Distribution Network #78581AK822695 · Hewlett-Packard Co. · Oct 15, 1982
- Intellivue Guardian Software Model 866009, Intellivue CL SP02 Pod Model 865215, Intellivue CL Nbp Pod Model 865216K111905 · Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · Oct 4, 2011
- Intellivue Clinical Information PortfolioK100272 · Philips Medical Systems · Apr 14, 2010
- Philips Drug Calculator SoftwareK023674 · Philips Medical Systems · Nov 29, 2002
- Metavision Clinical Information SystemK012349 · Imdsoft, Ltd. · Jan 29, 2002
| Jul 8, 2011 | Z-3111-2011Under certain circumstances, edits to an order can result in unintended scheduled interventions on the IntelliVue Clinical Information Portfolio. A patient may receive interventions or additional medication doses that… | Terminated |
| Mar 16, 2011 | Z-2175-2011Edits to an order can result in unintended scheduled interventions and a patient may receive additional interventions or medication doses that are not intended. | Terminated |
| Oct 19, 2010 | Z-0414-2011Software: Scheduling function of the Philips Intellivue Clinical Information Portfolio Releases D.00 through D.03. Edits to a scheduled even frequency order may result in pending interventions not being generated after… | Terminated |
| Oct 21, 2009 | Z-0418-2010A software defect exists in the ICIP Release 0.02 scheduling function for order management such that if a user edits either the Schedule Time or the Stop Time for an existing order, the corresponding intervention(s) for… | Terminated |
| May 5, 2009 | Z-1654-2009Error in therapy scheduling: If scheduled orders are documented and edited prior to the orders' original schedule time, ICIP can create a second instance of the same order marked as pending. A second instance of an… | Terminated |
| May 5, 2009 | Z-1436-2009Software: A software defect under certain circumstances could lead to the unintended administration of a duplicate dose of a scheduled medication within a single 24 hour period | Terminated |
| Mar 4, 2008 | Z-1541-2008, Z-1540-2008Medications prescribed for one patient were printed on the record of another patient. | Terminated |
- Nexxis 2;Nexxis OR 2;NMS 2;MNA-240;MNA-4x0 (Where “x” may be any numeric character).K261469 · Barco N.V. · Sep 30, 2026
- ESSENZ Patient MonitorK212003 · Livanova Deutschland, GmbH · Nov 10, 2021
- Philips IntelliVue GuardianSoftware (Rev. E.0X)K212208 · Philips Medizin Systeme Boeblingen GmbH · Sep 30, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Aug 6, 1999
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2450
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files