- Review time: 69 days
- Recalls: 4
- Cited as a predicate by: 3 devices
The Variant™ II β-thalassemia Program is intended for the separation and area percent determinations of hemoglobins A2 and F and as an aid in the identification of abnormal hemoglobins in whole blood using ion-exchange high performance liquid chromatography (HPLC).
- Premier Resolution SystemK222635 · Trinity Biotech (Primus Corporation, Dba Trinity Biotech) · Aug 4, 2023
- Modification To Variant Ii: Hemoglobin A1C Program, Beta-Thalassemia Short Program and Total Ghb ProgramK063643 · Bio-Rad Laboratories Inc., Clinical Systems Divisi · Dec 27, 2006
- Bio-Rad D-10 Dual ProgramK041444 · Bio-Rad Laboratories, Inc. · Jun 9, 2004
| Apr 8, 2014 | Z-1537-2014There is an issue with the buffers which impacts the quantification of the Hemoglobin F peak. Calibration and Quality control are likely to fail if the affected lots are used. | Terminated |
| Sep 12, 2011 | Z-0024-2012, Z-0025-2012, Z-0023-2012A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field… | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Apr 2, 1999
- Review panel
- Hematology
- Regulation
- 21 CFR 864.7400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files