Arca
Skip to content
Tools
Tools

Colorado Ii Spinal System

Orthosis, Spondylolisthesis Spinal Fixation. 510(k) K991031, from Sofamor Danek USA,Inc. (Memphis, TN). Product code MNH, class 2.

As of Nov 19, 2026, K991031 is an FDA 510(k) held by Sofamor Danek USA,Inc., substantially equivalent on May 10, 2000 after 408 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 408 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the COLORADO™ II Spinal System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal tumor, and/or (7) failed previous fusion…

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K991031, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Sep 10, 2007Z-0125-2008Assembly may be compromised: Two thread grooves are missing on the nut, which may cause damage to the bone screw and result in the need for the bone screw to be replaced.Terminated

Record

Type
Traditional
Received
Mar 29, 1999
Review panel
Orthopedic
Regulation
21 CFR 888.3070

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files