Medtronic PS Medical Introducer Sheath 12F,14F,15F, Models 3805-012, 3805-014, 3805-015

Endoscope, Neurological. 510(k) K990333, from Medtronic PS Medical (Goleta, CA). Product code GWG, class 2.

As of Nov 26, 2026, K990333 is an FDA 510(k) held by Medtronic PS Medical, substantially equivalent on Apr 28, 1999 after 84 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 84 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Introducer Sheath is a device used to make a channel through the brain into the ventricular system.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K990333, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jul 5, 2005Z-1111-05, Z-1110-05Cracked hubs/handles and/or improper peeling of the sheath during use.Terminated

Other GWG clearances

The newest 510(k)s for the same product code; their median review was 236 days; see every GWG device

Record

Type
Traditional
Received
Feb 3, 1999
Review panel
Neurology
Regulation
21 CFR 882.1480

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files