Arca
Skip to content

Bci Mini-Torr Plus, Model 6004

System, Measurement, Blood-Pressure, Non-Invasive. 510(k) K984618, from Bci Intl., Inc. (Waukesha, WI). Product code DXN, class 2.

As of Nov 17, 2026, K984618 is an FDA 510(k) held by Bci Intl., Inc., substantially equivalent on Mar 25, 1999 after 86 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Mar 25, 1999
  • Review time: 86 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The BCI Mini-Torr Plus monitor is a portable noninvasive blood pressure (NIBP) monitor for spot checking or monitoring of a patient's systolic, diastolic, and mean arterial (MAP) blood pressures, and pulse rate. Optional oximeter, electronic thermometer and/or integral printer are available.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K984618, newest first; each opens FDA's recall record
StartedReasonStatus
Apr 2, 2008Z-1649-2008Due to a component that may be installed backwards, the device may turn off or may immediately reset.Terminated

Other DXN Clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every DXN device

Record

Type
Traditional
Received
Dec 29, 1998
Review panel
Cardiovascular
Regulation
21 CFR 870.1130

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.