Set, Administration, Intravascular. 510(k) K984381, from Baxter Healthcare Corp (Round Lake, IL). Product code FPA, class 2.

As of Nov 20, 2026, K984381 is an FDA 510(k) held by Baxter Healthcare Corp, substantially equivalent on Feb 19, 1999 after 73 days, with 16 recalls tied to it in FDA's recall database.

  • Review time: 73 days
  • Recalls: 16
  • Cited as a predicate by: 2 devices

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K984381, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 7, 2012Z-0014-2013, Z-0013-2013, Z-0015-2013, Z-0012-2013Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.Terminated
Sep 3, 2010Z-0070-2011There are leaks in the Drip Chamber Check Valve Subassembly of the Solution Set. A leak in a sterile pathway could allow microbial entry, which could lead to a blood stream infection.Terminated
Nov 3, 2009Z-0487-2010The tubing may separate from the distal Clearlink Y-site of the set, which could result in patient injury if the separation occurs during patient use.Terminated
Dec 15, 2005Z-0655-06, Z-0654-06, Z-0656-06The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.Terminated
Mar 17, 2005Z-0199-06The Buretrol solution sets are currently labeled as acceptable for use with the Colleague Volumertric Pump, but there is a potential for a non-detection of an upstream occlusion under certain conditions.Terminated
Feb 28, 2005Z-0845-05Possible occlusion of the Clearlink injection site valve due to inadequate silicone lubrication of the slit in the upper seal of the valve.Terminated
Apr 28, 2004Z-0193-05Inadequate directions for use may have contributed to an increase in infection rates at a small number of customers beginning use of the Clearlink needleless device.Terminated
Oct 17, 2002Z-0508-03, Z-0510-03, Z-0523-03, Z-0504-03The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.Terminated

Other FPA clearances

The newest 510(k)s for the same product code; their median review was 266 days; see every FPA device

Record

Type
Traditional
Received
Dec 8, 1998
Review panel
General Hospital
Regulation
21 CFR 880.5440

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files