Arca
Skip to content
Tools
Tools

Leksell Gamma Knife Target System, Model 24001

System, Radiation Therapy, Radionuclide. 510(k) K984328, from Elekta Instrument AB (San Diego, CA). Product code IWB, class 2.

As of Nov 19, 2026, K984328 is an FDA 510(k) held by Elekta Instrument AB, substantially equivalent on May 21, 1999 after 169 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 169 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Leksell® Gamma Knife Target System Model 24001 is a teletherapy device indicated for use in the stereotactic irradiation of intracranial structures

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K984328, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 26, 2011Z-0312-2012Several of the LMR03 actuators with bronze drive nut have failed unexpectedly, creating a potential safety hazard for the operator and patient.Terminated
Oct 1, 2007Z-1236-2010After updating the LGK actuator in spare part 810361, the old sleigh became obsolete due to causing insufficient locking of the helmet in combination with the new actuator.Terminated
Sep 1, 2006Z-1533-2010Leksell Gamma Knife C 1.2 with MCU version 4.0.0.6 improperly adjusted trolley for 4mm helmet caused jamming when helmet changer attempted to lock helmet. Lock did not fully engage and could still be raised. When the…Terminated
Aug 1, 2006Z-1429-2010Due to a material change in a critical component of the linear actuators the quality of said part was found to be inadequate.Terminated

Other IWB clearances

The newest 510(k)s for the same product code; their median review was 135 days; see every IWB device

Record

Type
Traditional
Received
Dec 3, 1998
Review panel
Radiology
Regulation
21 CFR 892.5750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files