Angiographic Contrast Management System, Model Part A and Part B
Injector And Syringe, Angiographic. 510(k) K984231, from Invasatec (Eden Prairie, MN). Product code DXT, class 2.
As of Nov 20, 2026, K984231 is an FDA 510(k) held by Invasatec, substantially equivalent on May 31, 2000 after 553 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 553 days
- Recalls: 4
- Cited as a predicate by: 2 devices
The Acist™ CMS-2000 Angiographic Injection System with Contrast Management is intended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures.
- K963882K963882
- Fluid Administration Set - Contrast ManagerK961794 · Merit Medical Systems, Inc. · Aug 1, 1996
- Namic Contrast Savings Delivery SystemK903493 · North American Instrument Corp. · Mar 26, 1991
- ACISTCVi® Contrast Delivery SystemK171646 · Acist Medical Systems, Inc. · Feb 16, 2018
- C2000 Automated Manifold KitK052744 · Acist Medical Systems, Inc. · Dec 16, 2005
| Dec 21, 2015 | Z-0744-2016Reports related to ingress of air into the manifold. | Terminated |
| Oct 3, 2007 | Z-0359-2008Syringe Fracture: the syringe may be fractured, causing contrast to be sprayed from syringe fracture. | Terminated |
| May 19, 2004 | Z-1002-04Some of the sterile product packages were inadequately sealed prior to sterilization and as a result the package contents may not be sterile. | Terminated |
| Oct 2, 2003 | Z-0064-04Some of the automated manifold's 2.5 inch high-pressure lines for radiopaque contrast media, used in angiographic procedures, have separated where they are bonded to the manifold body of the device allowing contrast… | Terminated |
- MEDRAD® MRXperion MR Injection System and MR Injection System Syringe Kit; MEDRAD® IS12 Module (IS12); MEDRAD® Imaging…K262043 · Bayer Medical Care, Inc. · Oct 6, 2026
- Disposable Syringes with Accessories for Power Injectors (900103T); Disposable Syringes with Accessories for Power…K252652 · Fortis Biosolutions Company Limited · May 14, 2026
- VeraPro S100 Multi-Use Syringe (018808); VeraPro S100 LV1 Low Viscosity Single-Use Syringe (018811); VeraPro AMT…K252638 · Acist Medical Systems, Inc. · May 13, 2026
- ACIST Pro Diagnostic System (019304); Processing Unit Bed Mount; AngioTouch Module; ACIST Pro Cart; Single Transducer…K252653 · Acist Medical Systems, Inc. · May 13, 2026
- High Pressure Syringe; Pressure Connecting Tube; Quick fill tube; Spike; Transfer TubeK253048 · Shenzhen Maiwei Biotech Co., Ltd. · Apr 16, 2026
- Extension tubeK251788 · Zhuhai DR Medical Instruments Co., Ltd. · Mar 5, 2026
- MEDRAD MRXperion MR Injection System and MR Injection System Syringe Kit; MEDRAD ISI2 Module (ISI2)K252891 · Bayer Medical Care, Inc. · Feb 13, 2026
- High Pressure TubingK244038 · Shandong Int Medical Instruments Co., Ltd. · Sep 5, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 25, 1998
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files