Arca
Skip to content
Tools
Tools

Abbott Axsym Myoglobin

Myoglobin, Antigen, Antiserum, Control. 510(k) K983848, from Abbott Laboratories (Abbott Park, IL). Product code DDR, class 2.

As of Nov 23, 2026, K983848 is an FDA 510(k) held by Abbott Laboratories, substantially equivalent on Nov 16, 1998 after 17 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 17 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

AxSYM® Myoglobin is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative determination of myoglobin in human serum or plasma on the AxSYM System. Measurements obtained by this device aid in the rapid diagnosis of heart or renal disease.

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K983848, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 19, 2009Z-1050-2009Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.Terminated
Apr 23, 2007Z-1030-2007Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.Terminated

Other DDR clearances

The newest 510(k)s for the same product code; see every DDR device

Record

Type
Traditional
Received
Oct 30, 1998
Review panel
Immunology
Regulation
21 CFR 866.5680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files