Osteonics Secur-Fit Ad Generation Ii Acetabular Component System
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented. 510(k) K983502, from Osteonics Corp. (Allendale, NJ). Product code LPH, class 2.
As of Nov 17, 2026, K983502 is an FDA 510(k) held by Osteonics Corp., substantially equivalent on Dec 16, 1998 after 71 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Dec 16, 1998
- Review time: 71 days
- Recalls: 1
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Osteonics® Secur-Fit™-AD Generation II Acetabular Components are single-use devices. The shells are intended for cemented or cementless fixation within the prepared acetabulum. The Osteonics® Secur-Fit™-AD Generation II Acetabular Component System is compatible with any appropriately selected Osteonics hip stem/femoral head combination.
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- Osteonics Alumina C-Taper Head, Osteonics Trident Polyethylene InsertsK991952 · Howmedica Osteonics Corp. · Jun 29, 1999
- Omnifit® 20° Series I Insert;Omnifit® Crossfire® 10° Series II Insert; Constrained…K261624 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · Jul 10, 2026
- Trident X3 Acetabular Inserts, ADM and MDM X3 Acetabular Inserts, Accolade II Hip SystemK182468 · Howmedica Osteonics Corp. A.K.A. Stryker Orthopaedics · Mar 19, 2019
- Trident T Acetabular ShellsK040412 · Stryker Orthopaedics · May 25, 2004
- Trident Acetabular Shells: PsK001449 · Howmedica Osteonics Corp. · Aug 4, 2000
Recalls
Recalls FDA ties to K983502, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Jul 30, 2026 | Z-3204-2026Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in… | Open, Classified |
Other LPH Clearances
The newest 510(k)s for the same product code; their median review was 282 days; see every LPH device- G7® Acetabular SystemK261453 · Zimmer Biomet · Sep 23, 2026
- Mpact 3D Metal FoamK253630 · Medacta International S.A. · Aug 28, 2026
- TRIVANCE Hip StemK254043 · Globus Medical, Inc. · Aug 4, 2026
- Omnifit® 20° Series I Insert;Omnifit® Crossfire® 10° Series II Insert; Constrained…K261624 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · Jul 10, 2026
- iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with Plasma Sprayed CPTi…K243192 · Nextstep Arthopedix, LLC · Jun 5, 2026
- Taperloc® Complete Hip StemK251270 · Biomet, Inc. · May 19, 2026
- G7® Revision Acetabular SystemK252623 · Zimmer, Inc. · Feb 4, 2026
- Mpact 3D Metal Implants Extension DMLS TechnologyK251043 · Medacta International S.A. · Jan 28, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 6, 1998
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3358
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files