Arca
Skip to content
Tools
Tools

Bard Inlay Lubricious Double Pigtail Ureteral Stent With Suture (Hereinafter Referred To As Bard Lubricious Ureteral Ste

Stent, Ureteral. 510(k) K983498, from C.R. Bard, Inc. (Covington, GA). Product code FAD, class 2.

As of Nov 19, 2026, K983498 is an FDA 510(k) held by C.R. Bard, Inc., substantially equivalent on Dec 15, 1998 after 71 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 71 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Bard Lubricious Ureteral Stent is indicated to relieve obstruction in a variety of benign, malignant and post-traumatic conditions in the ureter such as presence of stones and/or stone fragments, or other ureteral obstructions such as those associated with ureteral stricture, carcinoma of abdominal organs, retroperitoneal fibrosis or ureteral trauma, or in association with Extracorporeal…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K983498, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Oct 13, 2016Z-0406-2017, Z-0404-2017Labeling: Statement was added to the Warnings/Precautions section requesting end users to consider the presence of a knot if significant resistance is encountered during attempts at removal.Terminated

Other FAD clearances

The newest 510(k)s for the same product code; their median review was 185 days; see every FAD device

Record

Type
Abbreviated
Received
Oct 5, 1998
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.4620

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files