Bio-Eye Hydroxyapatite Ocular Impland And Conformer
Implant, Eye Sphere. 510(k) K982562, from Integrated Orbital Implants, Inc. (San Diego, CA). Product code HPZ, class 2.
As of Nov 17, 2026, K982562 is an FDA 510(k) held by Integrated Orbital Implants, Inc., substantially equivalent on Dec 28, 1998 after 158 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Dec 28, 1998
- Review time: 158 days
- Recalls: 2
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Bio-eye Hydroxyapatite Ocular Implant is indicated as a primary implant in cases of enucleation and evisceration, and as a secondary implant in cases of poor performance of a primary implant, such as in cases of poor motility, migration, extrusion, chronic infection, enophthalmos, and lid sag.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Motility Orbital ImplantK891137 · Integrated Orbital Implants, Inc. · Aug 24, 1989
- Kolberg Ophthalmic ConformerK945110 · Kolberg Ocular Products, Inc. · Jan 13, 1995
- Interpore 200 Porous HydroxyapatiteK860983 · Interpore Intl. · May 23, 1986
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Ha Orbital ImplantK021643 · Laboratoire Villanova · Dec 13, 2002
Recalls
Recalls FDA ties to K982562, newest first; each opens FDA's recall record| Jun 7, 2010 | Z-2292-2010Integrated Orbital Implants, Inc. has initiated a recall of it's Sterile Bio-Eye Hydroxyapatite Orbital Implant and Conformer due to a labeling issue in which the expiration date of the product was omitted. | Terminated |
| Feb 27, 2006 | Z-0012-2007The end label on the outer box was mislabeled on 9 of a 10 unit lot. The end label says ''22 mm'' and the implants are actually ''20 mm''. The main label on the cardboard box applied by the contract packager is correct… | Terminated |
Other HPZ Clearances
The newest 510(k)s for the same product code; see every HPZ device- EzyporK212741 · Fci (France Chirurgie Instrumentation) Sas · Jun 24, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 23, 1998
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.3320
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files