Arca
Skip to content

Osteonics Scorpio Posterior Cruciate Retaining Total Knee System

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K974556, from Osteonics Corp. (Allendale, NJ). Product code JWH, class 2.

As of Nov 17, 2026, K974556 is an FDA 510(k) held by Osteonics Corp., substantially equivalent on Feb 27, 1998 after 84 days, with 7 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 27, 1998
  • Review time: 84 days
  • Recalls: 7
  • Cited as a predicate by: 11 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Osteonics® Scorpio™ Posterior Cruciate Retaining (CR) Total Knee System consists of posterior cruciate retaining total knee replacement femoral components, with corresponding cruciate retaining tibial bearing components, specifically designed to articulate together.

Cited as a Predicate

11 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K974556, newest first; each opens FDA's recall record
Aug 15, 2014Z-1634-2015The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.Terminated
Oct 2, 2009Z-0466-2010It was determined that a change made in the dimensions of the collar component of the packaging resulted in the inner blister material being overstretched causing a potential hole in the sterile barrier system.Terminated
Oct 2, 2009Z-0467-2010It was determined that a change made in the dimensions of the collar component of the packaging resulted in the inner blister material being overstretched causing a potential hole in the sterile barrier system.Terminated
Oct 2, 2009Z-0468-2010It was determined that a change made in the dimensions of the collar component of the packaging resulted in the inner blister material being overstretched causing a potential hole in the sterile barrier system.Terminated
Aug 11, 2005Z-0059-06The Left Scorpio Total Knee Cruciate Retaining Waffle Femoral Component #7 packaged and labeled as left may contain a Right and vice versa.Terminated
Aug 11, 2005Z-0058-06The Left Scorpio Total Knee Cruciate Retaining Waffle Femoral Component #7 packaged and labeled as left may contain a Right and vice versa.Terminated
Feb 8, 2005Z-1381-2008Mislabeled: The labeling indicates a Scorpio CR Femoral Component, however, the device inside the package is actually a Scorpio PS Femoral Component.Terminated

Other JWH Clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Dec 5, 1997
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files