Arca
Skip to content
Tools
Tools

Ohmeda-Ohio Care Plus Incubator

Incubator, Neonatal. 510(k) K974349, from Ohmeda Medical (Columbia, MD). Product code FMZ, class 2.

As of Nov 25, 2026, K974349 is an FDA 510(k) held by Ohmeda Medical, substantially equivalent on Feb 12, 1998 after 85 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 85 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. Most incubators can be used in two operating modes: 1.

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K974349, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 26, 2023Z-0957-2024GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed…Open
Nov 18, 2016Z-1510-2017Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke,…Terminated

Other FMZ clearances

The newest 510(k)s for the same product code; their median review was 179 days; see every FMZ device

Record

Type
Traditional
Received
Nov 19, 1997
Review panel
General Hospital
Regulation
21 CFR 880.5400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.