Arca
Skip to content
Tools
Tools

Versys Hip System Beaded Hip Prosthesis

Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented. 510(k) K973714, from Zimmer, Inc. (Warsaw, IN). Product code LPH, class 2.

As of Nov 19, 2026, K973714 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Dec 24, 1997 after 86 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 86 days
  • Recalls: 8
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The VerSys Beaded Hip Prosthesis is designed to achieve biologic fixation to bone and is indicated for: Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur, congenital hip dysplasia, protrusio…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K973714, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 29, 2017Z-1024-2018, Z-1023-2018, Z-1025-2018, Z-1026-2018, Z-1022-2018The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.Terminated
Aug 1, 2017Z-0722-2018The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.Terminated
Jan 11, 2016Z-0848-2016LDPE bag containing the implant adheres to the highly polished implant surface.Terminated
May 19, 2014Z-1842-2014During routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively.Terminated

Other LPH clearances

The newest 510(k)s for the same product code; their median review was 282 days; see every LPH device

Record

Type
Traditional
Received
Sep 29, 1997
Review panel
Orthopedic
Regulation
21 CFR 888.3358

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files