Leksell Gammaplan
System, Radiation Therapy, Radionuclide. 510(k) K973441, from Patterson Consulting Group, Inc. (San Diego, CA). Product code IWB, class 2.
As of Nov 19, 2026, K973441 is an FDA 510(k) held by Patterson Consulting Group, Inc., substantially equivalent on Apr 1, 1998 after 202 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 202 days
- Recalls: 4
- Cited as a predicate by: 4 devices
The Leksell GammaPlan (LGP) is designed for use with the Leksell Gamma Knife manufactured by Elekta Instrument, AB. The Leksell GammaPlan is intended to be used for planning the dosimetry of treatments in stereotactic radiosurgery and stereotactic radiation therapy.
- Leksell Gammaplan (Lgp)K914808 · Elekta Instruments, Inc. · Jan 23, 1992
- Leksell GammaplanK090972 · Elekta Instrument AB · Jul 20, 2009
- Leksell GammaplanK061540 · Elekta Instrument AB · Jul 5, 2006
- GammaplanK051022 · Elekta Instrument AB · Jun 1, 2005
- Leksell Gammaplan 4C With MultiviewK042269 · Elekta Instrument AB · Sep 8, 2004
| Jun 23, 2011 | Z-0307-2012There is a risk that the user may interpret the instructions in the manual too literally, and adjust the TMR 10 dose description to obtain an ETCPD that matches previous clinical prescriptions, which in turn could lead… | Terminated |
| Oct 23, 2008 | Z-1134-2010Although the co-registration looks good during the verification step in the co-registration dialog, the obtain transformation may include an error that depends on the voxel sizes and acquisition parameters of the… | Terminated |
| Jan 7, 2008 | Z-1604-2010The precision of the calculation used to create the fused study in LGP 8.0 is too low and should not be used until the system is upgraded to LGP 8.2. | Terminated |
| Oct 1, 2005 | Z-1223-2011Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped. | Terminated |
- Leksell Gamma Knife® (Elekta Esprit); Leksell Gamma Knife® (Icon™); Leksell Gamma Knife® (Perfexion™)K250391 · Elekta Solutions AB · Jul 2, 2025
- Leksell GammaPlan (LGP)K232854 · Elekta Solutions AB · Feb 8, 2024
- Leksell Gamma Knife Perfexion, Leksell Gamma Knife Icon, Leksell Gamma Knife - Elekta Esprit (New model variant)K222047 · Elekta Solutions AB · Oct 26, 2022
- TaiChiCK203250 · Our United Corporation · Mar 12, 2021
- Akesis Galaxy RTiK203146 · Akesis, Inc. · Mar 5, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 11, 1997
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files