Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K973412, from Sulzer Orthopedics, Inc. (Austin, TX). Product code JWH, class 2.

As of Nov 17, 2026, K973412 is an FDA 510(k) held by Sulzer Orthopedics, Inc., substantially equivalent on Dec 8, 1997 after 90 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 8, 1997
  • Review time: 90 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Natural-Knee™ II Constrained Knee System is intended use in the following diagnostic indications: 1. Patient conditions, including but not limited to, inflammatory degenerative joint disease (e.g., rheumatoid arthritis) and noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis). 2. Correctable valgus-varus deformity and moderate flexion contracture. 3.

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K973412, newest first; each opens FDA's recall record
Oct 2, 2014Z-1948-2015Failed to meet established requirements for sterile barrier integrity due to a breach of the inner cavity. If the outer tray is compromised in addition to the inner tray a periprosthetic infection may occur. The risk…Terminated
Feb 25, 2014Z-1256-2014A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the…Terminated

Other JWH Clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Sep 9, 1997
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.