Arca
Skip to content
Tools
Tools

Osteonics Scorpio Total Knee Universal Dome Patellar Component and 1 more

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K972967, from Osteonics Corp. (Allendale, NJ). Product code JWH, class 2.

As of Nov 18, 2026, K972967 is an FDA 510(k) held by Osteonics Corp., substantially equivalent on Jan 16, 1998 after 158 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 158 days
  • Recalls: 6
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Osteonics® Scorpio™ Total Knee Universal Dome and Recessed Patellar Components are intended to articulate with any commercially available Osteonics® Scorpio™ Total Knee Femoral Component, any commercially available Osteonics® Series 7000 Total Knee Femoral Component, and any commercially available Osteonics® Omnifit® Total Knee Femoral Component.

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K972967, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Dec 1, 2023Z-0893-2024, Z-0895-2024, Z-0894-2024, Z-0900-2024, Z-0896-2024, Z-0897-2024Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.Open

Other JWH clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Aug 11, 1997
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.