Arca
Skip to content
Tools
Tools

Olympus GF-UM130 Ultrasound Gastrovideoscope, Its Associated Accessories and Ancillary Equipment

Gastroscope And Accessories, Flexible/Rigid. 510(k) K971660, from Olympus America, Inc. (Melville, NY). Product code FDS, class 2.

As of Nov 18, 2026, K971660 is an FDA 510(k) held by Olympus America, Inc., substantially equivalent on Jul 25, 1997 after 80 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 80 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Olympus GF-UM130 Ultrasound Gastroscope has been designed to be used with an Olympus Endoscopic Ultrasound Center, EVIS Video System Center, light source, documentation equipment, video monitor, Endo-Therapy accessories and other ancillary equipment for endoscopic ultrasound imaging of the gastrointestinal wall, biliary and pancreatic duct, and surrounding organs.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K971660, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Oct 16, 2020Z-0502-2021, Z-0503-2021A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.Terminated

Other FDS clearances

The newest 510(k)s for the same product code; their median review was 167 days; see every FDS device

Record

Type
Traditional
Received
May 6, 1997
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.