Servo Ventilator 300A
Ventilator, Continuous, Facility Use. 510(k) K970839, from Siemens Elema AB (S-17195 Solna, SE). Product code CBK, class 2.
As of Nov 17, 2026, K970839 is an FDA 510(k) held by Siemens Elema AB, substantially equivalent - subject to tracking reg. on Oct 14, 1997 after 221 days, with 4 recalls tied to it in FDA's recall database.
- Substantially Equivalent - Subject to Tracking Reg.: Oct 14, 1997
- Review time: 221 days
- Recalls: 4
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFUse of the Servo Ventilator 300A is indicated for adult, pediatric or neonatal patient populations in an environment where patient care is provided by Healthcare Professionals (Physician, Nurse, Technician), when the professional determines that a device is required to assist the breathing of the patient.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Servo Ventilator 300 Computer InterfaceK960010 · Siemens Elema AB · Oct 25, 1996
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- Carina HomeK060705 · Draeger Medical B.V. · Jun 14, 2006
- Esprit Ventilator Neonatal Option, Model V1000K051262 · Respironics California, Inc. · Jul 8, 2005
- Servo-I Ventilator System, Model 64 87 800 E407EK040221 · Maquet Critical Care AB · Feb 20, 2004
- Siemens Kion Anesthesia Workstation - Extended Modes Functionality, Models: 64 62 910 E392E, 65 22 572 E397E, 65 03 655K024213 · Siemens Medical Solutions USA, Inc. · Mar 23, 2004
Recalls
Recalls FDA ties to K970839, newest first; each opens FDA's recall record| Nov 17, 2009 | Z-0588-2010Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the… | Terminated |
| Mar 23, 2009 | Z-1252-2009The expiratory valve in the SV 300 ventilator may stay in the open position if the printed circuit board fails. | Terminated |
| Dec 19, 2008 | Z-0872-2009SERVO Ventilator 300A could cause hypoventilation if the use of the PC or PRVC mode with the Automode function set to "On" in Neonatal and Pediatric patient ranges with small tidal volumes and short set inspiratory rise… | Terminated |
| Jan 3, 2005 | Z-0571-05Failures of the air and oxygen gas modules occurred because of misuse/non-compliance of directions. | Terminated |
Other CBK Clearances
The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits Size A (ONIV117A); F&P OptiNIV…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
Record
- Type
- Traditional
- Received
- Mar 7, 1997
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files